General

Operating Temperature Range

-20 to 60 °C (-5 to 140 °F)

Storage Temperature Range

-30 to 60 °C (-22 to 140 °F)

Relative Humidity Range

10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing

Size

L=5.63 x W=3.00 x H=1.60 inches

Weight

12.1 ounces (including boot & batteries)

Batteries

Four “AA” Alkaline 1.5V (LR6)

Optional AC Adaptors

120 VAC 50/60 Hz [Part # 020-0100]
240 VAC 50/60 Hz [Part # 020-0101]

Optional NiMh Rechargeable battery kit

120 VAC for North America Only; charger,  four NiMh 

batteries, AC & DC cords [Part # 020-0103]

Low Battery

Low battery indication with nominal 1 hour of operation 

left

Protection against misconnection

Over-voltage protection to 135 vrms (rated for 30 seconds) 

or 240 vrms (rated for 15 seconds)

Display

High contrast backlit graphic liquid crystal display with 

0.413” (10.5 mm) high digits

Read mA

334 Ranges and Resolution

0.00 to 52.00 mA Full Span OR -25.0 to 300.0% of 4-20 mA

Accuracy

Below 24.01mA 
Above 24.00mA

≤ ± (0.05 % of 24.00mA)   (± 0.01mA)
≤ ± (0.05 % of 52.00mA)   (± 0.03mA)

Voltage burden

≤ 2V at 50 mA

Overload/Current limit protection

54 mA nominal

Battery life

≥ 125 Hours nominal

PIECAL 334 Specifications

(Unless otherwise indicated all specifications are rated from a nominal 23 °C,

70 % RH for 1 year from calibration)

Practical Instrument Electronics, Inc.

Available at www.Instrumentation.com 

Summary of Contents for piecal 334

Page 1: ...to check loop response and control settings The PIECAL Model 334 uses any loop power from 2 to 100V DC READ LOOP CURRENT Check controller outputs or measure the milliamp signal anywhere in the loop T...

Page 2: ...displayed in the units selected r DIAL KNOB Turn the knob to adjust output level Turn clockwise to increase the output counter clockwise to decrease the output t EXTERNAL POWER JACK Not Shown When us...

Page 3: ...while the EZ CHECK switch e is in the DIAL position or the current can be set at the xed points of 4 00mA 0 0 or 20 00mA 100 0 with switch e READ mA READ Percent of 4 to 20 mA Choose this function to...

Page 4: ...IECAL Model 334 to the plus input of the device and the black source lead to the minus The PIECAL Model 334 supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter The current passed by the...

Page 5: ...RCE output to full scale supplies a nominal 24V DC to power a 2 Wire Transmitter while simultaneously displaying the 4 to 20 mA output of the transmitter READ MILLIAMPS Select READ milliamps by moving...

Page 6: ...Electronics Inc be liable for any special incidental or consequential damage SETTING UP VALVES When setting up a valve it is important to correctly set the end stops Use the PIECAL Model 334 wo supply...

Page 7: ...dication with nominal 1 hour of operation left Protection against misconnection Over voltage protection to 135 vrms rated for 30 seconds or 240 vrms rated for 15 seconds Display High contrast backlit...

Page 8: ...V at 20 mA Overload Current limit protection 24 mA nominal Loop voltage limits 2 to 100 VDC fuse less protected from reverse polarity connections Battery life 125 hours nominal Voltage Read Range and...

Reviews: