SOURCE

Choose this function to provide a simulated RTD signal into controllers, temperature transmitters, 
indicators or any input devices that measure thermocouple sensors.

1) Disconnect the RTD sensor from the device to 

be calibrated.

2)  Select “

SOURCE

” with slide switch 

w

.

3)  Connect the PIECAL 211 to the device using 2, 

3 or 4 wires matching the connections of the 
sensor that was just removed.

The output is adjusted in 0.1° (or 0.01/0.1 ohm) 
increments by turning the knob 

e

 

while the 

EZ-CHECK™ switch 

q

 is in the “HI”, “LO” or “SET”  

position. Press and turn the knob for faster dialing 
with 10° (or 1.00/10.0 ohm) increments.

The OVERLOAD indicator will light if excessive 
voltage or current is detected by the calibrator.

Calibrating RTD Instruments

3 2 2 - 1

THERMOCOUPLE

CALIBRATOR

J, T, E, K, mV

HI

Pt100 

α

=385

AUTOMATED

RTD

CALIBRATOR

READ

SOURCE

HI

SET

LO

MAX
READ

MIN

OVERLOAD

Push & Hold

to

Store/Step

Push & Turn

for

Fast Dialing

Double Click

for Options

OFF

°C

3W   2W      1W  4W

     

          

    

100.0

READ

Choose this function to measure temperatures with an RTD probe or sensor.

1)  Disconnect the RTD sensor from any other device.
2)  Select “

READ

” with slide switch 

w

.

3)  Place the EZ-CHECK switch into the READ position.
4)  Connect the PIECAL 211 to the device using 2, 3 or 4 wires.

The PIECAL 211 measures the temperature signal and constantly updates the display with the 
current temperature reading. Move the EZ-CHECK switch to MAX to see the highest temperature 
reading and to MIN to see the lowest temperature reading. Press and hold the knob 

e

 to clear the 

MAX and MIN readings.

The OVERLOAD indicator will light if excessive voltage or current is detected by the calibrator.

Reading RTD Sensors

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for PIECAL 211

Page 1: ...s for instant recall with the EZ CHECK switch Calibrate quickly with automatic output stepping Choose between 2 3 5 11 and 21 steps to automatically increment the output in 100 50 25 10 or 5 of span Select the step time to match your system from 5 6 7 8 10 15 20 25 30 and 60 seconds Compatible with all process instruments Connect directly to the RTD inputs of smart transmitters PLCs DCS and multic...

Page 2: ...k HI LO points in SIMULATE mode Continue to press and hold the knob for two more seconds to start the automatic ramping READ Press and hold to transfer the current temperature into the EZ Check HI LO points This clears the HI LO temperature readings which will update as the temperature changes Double click the knob to get into the PIECAL 211 Configuration Mode Use configuration to select C or F T ...

Page 3: ...ob will toggle between C and F RTD pressing the knob will cycle through the various RTD Pt Cu Ni at different base resistances and alpha values BACKLIGHT If BACKLIGHT is ON the backlight will light all the time the unit is powered up For maximum battery life turn the backlight off when using the calibrator in areas with enough ambient light to read the display STEPS pressing the knob will cycle th...

Page 4: ... the calibrator Calibrating RTD Instruments 3 2 2 1 THERMOCOUPLE CALIBRATOR J T E K mV HI Pt100 α 385 AUTOMATED RTD CALIBRATOR READ SOURCE HI SET LO MAX READ MIN OVERLOAD Push Hold to Store Step Push Turn for Fast Dialing Double Click for Options OFF C 3W 2W 1W 4W 100 0 READ Choose this function to measure temperatures with an RTD probe or sensor 1 Disconnect the RTD sensor from any other device 2...

Page 5: ...cing the EZ CHECK Switch into the HI or LO position then press and hold the e DIAL KNOB for 6 seconds the word STORE will appear on the display after 3 seconds and continue to press the DIAL KNOB until the word STEPPING appears on the display The word STEPPING will appear on the display anytime the selected automatic function is running Stop the stepping by again pressing and holding the e DIAL KN...

Page 6: ...328 0 to 392 0 392 0 to 1200 0 5 0 7 Pt 100 Ohm US Lab 1 3926 0 003926 200 0 to 100 0 100 0 to 700 0 700 0 to 862 2 0 3 0 4 0 5 328 0 to 212 0 212 0 to 1292 0 1292 0 to 1584 0 0 5 0 7 0 8 Pt 1000 Ohm 1 3850 0 00385 200 0 to 200 0 200 0 to 266 0 0 3 0 4 328 0 to 392 0 392 0 to 511 0 0 5 0 7 Pt 1000 Ohm Hy Cal HVAC 1 3750 0 00375 200 0 to 200 0 200 0 to 274 0 0 3 0 4 328 0 to 392 0 392 0 to 525 0 0 ...

Page 7: ...nst misconnection Over voltage protection to 60 V dc rated for 30 seconds Display High contrast graphic liquid crystal display LED backlighting for use in low lit areas Read Excitation Current 0 5 mA nominal Normal Mode Rejection 50 60 Hz 50 dB Common Mode Rejection 50 60 Hz 120 dB PIECAL 211 Specifications Unless otherwise indicated all specifications are rated from a nominal 23 C 70 RH for 1 yea...

Page 8: ...ondition shall Practical Instrument Electronics Inc be liable for any special incidental or consequential damage Additional Information This product is calibrated on equipment traceable to NIST and includes a Certificate of Calibration Test Data is available for an additional charge Practical Instrument Electronics recommends a calibration interval of one year Contact your local representative for...

Reviews: