6
HOW TO SWITCH THE BLOOD GLUCOSE METER ON AND OFF
The glucose meter is switched on by one of the following methods:
• Pressing the
C
button once;
• Pressing the down scroll button (
▼
) once;
• Press and hold the up scroll button (
▲
) and the down scroll button (
▼
) at the same time.
• Inserting the reactive strip in the test strip slot.
The glucose meter is switched off by one of the following methods:
• Pressing the
C
button for 3 seconds;
• Automatically after 3 minutes of idle time;
• Removing the reactive strip with test strip ejector.
THE FOUR MEASURING MODES
The meter provides you with four modes for measuring, General (no show), AC, PC and QC. You can switch
between each mode by:
• Start with the meter switched off. Insert a test strip to turn on the meter. The screen will display “CHK” and a
flashing “ ” .
• Press
C
to switch between General (no show), AC, PC and QC mode.
QUALITY CONTROL TESTING
When Should the Control Solution Test be Performed?
• if it is mandatory following the local regulations in your country,
• if you suspect the meter or test strips are not working properly,
• if your blood glucose test results are not consistent with how you feel, or if you think the results are not accurate,
• to practice the testing process, or
• if you have dropped or think you may have damaged the meter.
The items specified below may not be included in the kit (please check the contents on your product box) but
are required for control solution test or blood glucose test. Please make sure you have those items needed for a
test beforehand: Test strips
(c),
control solutions
(d)
, lancing device
(e)
or sterile lancets
(f)
. Please contact the
consumers’ toll-free number for Italy 800 900 080 and for France 01.55.93.26.48. to request the control solution.
Performing a Control Solution Test
To perform a control solution test, you will need:
(b)
,
(c)
and
(d)
.
1.
Insert the test strip to turn on the meter
Insert the test strip into the meter. Wait for the meter to display “CHK”,and a flashing “ ”.
2.
Press C to mark this test as a control solution test
With “
QC
” displayed, the meter will store your test result in memory under “
QC
”. If you press C again, the “
QC
”
will disappear and this test is no longer a control solution test.