PI PiezoMove P-601 Short Instructions Download Page 5

M O T I O N   |   P O S I T I O N I N G

PZ302EK 1.0.0 11/2019. 0.5

Gluing the piezo actuator on

1.  Glue the dowel pins into the locating holes 

of the piezo actuator‘s base body or into 

the equipment intended for mounting the 

piezo actuator.

2.  Align the piezo actuator and the equipment 

so that each dowel pin fits in the correspon-

ding locating hole.

3.  Glue the piezo actuator to the equipment.

4.  Wait until the adhesive has hardened 

completely.

5.  Check for firm seating of the piezo actuator 

in the equipment.

Checking the Connection of the Protec­

tive Earth Conductor

 

X

Make sure that the contact resistance at 

all protective earth connections is <0.1 Ω 

at 25 A.

 

X

If ground loops occur, contact the PI custo-

mer service department.

Screwing the Load

Tools and accessories

Suitable fixing screw(s) (depending on the 

model; refer to user manual)

Suitable tool to hold the movable part 

firmly

Screwdriver or wrench

Requirements

You have fixed the piezo actuator properly.

The piezo actuator is not connected to the 

electronics.

Models with stranded wires: The piezo 

actuator is discharged and short-circuited.

The distance between the center of gravity 

of the load and the center of the movable 

part is as small as possible in all directions.

Screwing the load

1.  Align the load so that the mounting holes 

(or mounting hole) in the movable part of 

the piezo actuator can be used to fix the 

load.

2.  Hold the movable part of the piezo actuator 

firmly.

3.  Fix the load to the movable part of the 

piezo actuator. Do not exceed the respec-

tive maximum permissible torque for the 

movable part and the screws used (refer to 

user manual).

4.  Loosen the movable part of the piezo 

actuator that was held firmly so it can move 

freely again.

The instructions required for startup and 

operation are in the manual for the electronics 

used.

The piezo actuator must be discharged in the 

following cases:

Before short-circuiting

When not in use

Before reconnecting if the connecting cable 

is accidentally pulled out of the electronics 

during operation

Before assembly work

Discharging the Piezo Actuator Connec­

ted to the Electronics

 

X

Set the piezo voltage on the electronics to 

0 V.

Discharging the Piezo Actuator Not 

Connected to the Electronics

Piezo actuator without connector:

1.  Ensure adequate protection against 

touching live parts.

2.  Short-circuit the piezo actuator for at 

least a few seconds with a suitable 10 kΩ 

discharge resistor.

Piezo actuator with connector: 

 

X

Connect the piezo actuator's voltage plug 

to the switched-off electronics from PI for 

at least a few seconds.

Short­Circuiting a Discharged Piezo 

Actuator

1.  Disconnect the piezo actuator from the 

electronics if it is still connected.

2.  Short-circuit the piezo actuator by twisting 

the stranded wires, using a suitable shor-

ting clamp or a suitable shorting plug.

Starting and Operating the 

Piezo Actuator

Discharging and Short­

Circuiting the Piezo 

Actuator