Intended Use

The piezo actuator is a laboratory device as 

defined by DIN EN 61010-1. It is intended for 

indoor use and use in an environment that is 

free of dirt, oil, and lubricants.
In accordance with its design, the piezo actua-

tor is intended for integration into a mechanical 

system for the positioning of loads, dynamic 

positioning, vibration absorption, and force 

generation. The operator is responsible for 

a standards-compliant integration into the 

overall system.
The piezo actuator is not intended for applica-

tions in areas where failure would be a consi-

derable risk for people or the environment.
The motion of the piezo actuator is along one 

axis.
The intended use of the piezo actuator is 

only possible when completely mounted and 

connected. The piezo actuator must be ope-

rated with suitable electronics (refer to user 

manual). The electronics are not included in the 

scope of delivery of the piezo actuator.
The piezo actuator may not be used for 

purposes other than those stated in the user 

manual. The piezo actuator may only be used 

in compliance with the technical specifications 

and instructions in the user manual.

Electrical Dangers

Temperature changes and compressive stress 

can induce charges in piezo actuators. After 

disconnecting from the electronics, the piezo 

actuator can remain charged for several hours. 

Touching live parts can result in minor injury 

from electric shock.

 

X

Do not disassemble the piezoelectric 

actuator.

 

X

Do not touch any live parts (depending on 

the model, e.g., open piezo actuator, open 

ends of stranded wires, or contacts in the 

plug connector).

 

X

Remove the shorting clamp or shorting plug 

that is in the scope of delivery only if neces-

sary for installation or operation purposes.

 

X

If you disconnect the piezo actuator from 

the electronics, discharge it and keep it 

short-circuited after discharging.

 

X

Do not disconnect the piezo actuator from 

the electronics during operation.

If the protective earth conductor is missing or 

not properly connected, touching the piezo 

actuator can result in minor injuries from 

electric shock in the case of a malfunction.

 

X

Operate the piezo actuator only with a 

properly connected protective earth 

conductor.

 

X

Observe the applicable standards for moun-

ting the protective earth conductor.

Electric flashovers can be caused by moisture, 

liquids, and conductive materials (e.g., metal 

dust) and can destroy the piezo actuator during 

operation. In addition, electric flashovers can 

also occur in certain air pressure ranges due to 

the increased conductivity of the air.

 

X

Avoid operating the piezo actuator in 

environments that can increase the electric 

conductivity.

 

X

Operate the piezo actuator only under 

permissible ambient conditions (refer to 

user manual).