2. Calculate the peak current requirement for sinusoidal operation using the following
formula:
Variable Description
Notes
I
a
Required average current of the ampli-
fier (source / sink) [A]
It is essential that the power adapter
can supply enough current.
I
max
Required peak current of the amplifier
(source / sink) [A]
The provided peak current depends on
the internal storage capacity of the am-
plifier.
f
Operating frequency [Hz]
Details on the operating frequency see
"Overview of Limiting Factors" (p. 18)
.
C
Capacitance of the piezo actuator [F (=
As/V)]
See
for the small-
signal capacitance of the piezo actua-
tor.
For large-signal conditions, a safety
factor of 70 % should be added to the
small-signal capacitance.
U
p-p
Operating voltage (peak-to-peak) [V]
Voltage difference between positive
and negative peak voltage
7.2
Starting and Operating the P-843
Tools and Accessories
■
Requirements
✓
You have read and understood the
general safety instructions (p. 8)
✓
You have
the P-843 correctly.
✓
You have read and understood the user manual for the electronics used.
✓
If a digital controller is used: You have read and understood the manual for the PC
software used.
✓
The electronics and if required, the PC software, have been installed (see the user manual
for the electronics).
CAUTION
Risk of electric shock if the protective earth conductor is not connected!
If the protective earth conductor is missing or not properly connected, risk of dangerous
touch voltages on the P-843 in the event of malfunction or failure of the system. If there are
touch voltages, touching the P-843 can lead to minor injury due to electric shock.
►
Connect the P-843 to a protective earth conductor before startup.
►
Do not remove the protective earth conductor during operation.
►
If the protective earth conductor has to be removed temporarily (e.g., for modifications),
reconnect the P-843 to the protective earth conductor before restarting.
7 STARTUP AND OPERATION
P843M0001EN
‒
2/23/2021
20