7 Start-Up
42
Version: 2.0.0
MS205Equ
C-863 Mercury Controller
7.2 Adapting the DIP Switch Settings
7.2.1 General Procedure
INFORMATION
Changed DIP switch settings become effective after the C-863 is switched on.
If you have changed the DIP switch settings while the C-863 was switched on,
switch the C-863 off and back on again to activate the new settings.
Figure 7: DIP switches: switch up = ON; switch down = OFF
Switches
Function
1 to 4
Controller address (p. 43); 16 possible combinations
5 and 6
Baud rate (p. 44)
7
Logic level of the limit switches (p. 44)
8
Update mode (p. 45)
Prerequisite
The C-863 is switched off, i.e. the power supply is
not
connected to the power
socket with the power cord.
Adapting the DIP switch settings
Put the individual DIP switches in the correct position for your application.
Details are given in the following tables.