background image

 

 

www.phywe.com, © All rights reserved 

11260-

02 / -88 / 2516

 

4.1  Putting the ripple tank into operation 

First connect the connecting socket (

11

of the ripple tank via 

the  provided  power  supply  to  the  mains.  The  LCD  screen 
lights up and the ripple tank is ready to use. 

4.2  Filling the wave tray 

First,  the  empty  wave  tray  has  to  be  placed  on  the  ripple 
tank.  Afterwards,  it  can  be  carefully  filled  with  water.  The 
wash-bottle supplied can be used to fill the wave tray. Fill the 
tray about three quarters high. After this, to have reflection of 
water  waves  at  the  sides  suppressed,  moisture  the  whole 
side-foam with a finger so that water can wet it. 
Use the adjusting screws (

14

) to align the wave tray horizon-

tally.  To  do  so,  we  recommend  that  you  look  flat  across  the 
water surface and use the height of the water level at the dif-
ferent sides of the tray as reference. Now use the screws to 
bring the tray to about the same height on all sides and so to 
a horizontal position. 

 

When  lifting  the  filled  tray,  the  water  tends  to  slop!  Pay 
attention that no water enters the device. 

We  recommend  using  the  wash-bottle  or  e.g.  a  sponge  to 
finally remove the water while the wave tray is still resting on 
the ripple tank. 

4.3  Cleaning 

The cover plate and the wave tray of the ripple tank consist of 
acrylic glass. Please perform cleaning of the cover plate and 
the wave tray with suitable (i.e., non-abrasive) tissues without 
aggressive chemicals added. 

4.4  Generation of circular waves 

Fix  the  holding  rod  (

16

)  with  a  single  plug  (

17

)  to  the  inte-

grated  generator  (

15

)  and  bring  it  to  the  middle  of  the  wave 

tray. Turn the clamping screw on the generator until the plug 
is  dipped  into  the  water  in  the  wave  tray.  Set  to  continuous 
light, to a frequency of between 5

 Hz and about 20 Hz and to 

a  small  amplitude  (1  or  2).  Prepare  the  observation  surface 
by  placing  a  sheet  of  white  paper  on  the  drawing  table  situ-
ated over the wave tray. Darkening the room could be useful 
for better visibility of the wave image. 
Should the wave image be distorted or unclear, vary the fre-
quency 

f

 and/or the amplitude until a wave image is obtained 

that is similar to that shown in Fig. 7. 

 

Fig. 7:   Circular waves generated by the plug-shaped exciter. 

4.5  Generation of flat waves 

The  generation  of  flat  waves  is  somewhat  more  difficult,  be-
cause it requires an exact alignment of the wave tray with the 
slide  used.  It  could  be  necessary  to  add  1  to  2  drops  of  a 
soap  solution  to  the  water  for  better  wetting.  Do  not  exceed 
this  amount,  however,  as  then  turbidity  and  foaming  would 
occur and cause deterioration in the quality of the image. 

Fix the holding rod with slide to the exciter and bring it to the 
lower end of the wave tray. The wave tray must now be cor-
rectly adjusted, as the slide must be exactly evenly immersed 
into  the  water.  If  this  is  not  done  precisely,  the  wave  image 
will  be  distorted  and  unclear.  See  the  information  given  on 
the correct adjustment of the wave tray in section 4.2 

“Filling 

the wave tray

The  slide  must  also  be  set  to  be  horizontal.  Therefore,  fix  it 
as  horizontally  as  possible  to  the  integrated  exciter  and 
tighten  the  clamping  screw  on  the  exciter  further  until  the 
slide  is  held  slightly  above  the  water  surface.  Use  the  water 
surface as reference to adjust the slide to the horizontal posi-
tion.  Afterwards,  turn  the  clamping  screw  on  the  exciter  fur-
ther until the slide is dipped into the water (abou

t 1 to 2 mm). 

Set  a  frequency  between  18

 Hz  and  25 Hz  and  choose  a 

small  amplitude  as  well  as  continuous  light.  A  wave  image 
should be obtained that is similar to that shown in Fig. 8. 

 

Fig. 8:  Flat waves generated by the plane exciter. 

Should the wave image be distorted or unclear, vary the fre-
quency, the amplitude or the immersion depth. When no im-
provement  in  the  wave  image  results,  it  could  be  possible 
that the slide is not correctly adjusted and therefore does not 
dip  evenly  into  the  water.  In  this case,  align  the  slide  as  de-
scribed above. 

4.6  Usage of the stroboscopic light 

The use of stroboscopic light is only  instructed in those indi-
vidual  experiments  in  which  it  is  required  to  reach  the  ex-
perimental  target.  All  other  experiments  should  first  be  car-
ried  out  without  stroboscopic  light  because stroboscopic illu-
mination considerably  changes  the  impression of  the  image, 
partially.  Interference  patterns  can  be  far  more  impressively 
seen  when  continually  projected  than  when  under  strobo-
scopic illumination. 
In this context it must be considered that for continuous light, 
the  human  eye  is  only  capable  of  recognizing  the  wave  im-
age  in  the  lower  frequency  range.  Because  of  this,  we  rec-
ommend that only frequencies of up to about 30

 Hz be used 

under  continuous  light.  Stroboscopic  light  should  be  used 
with  shorter  wavelengths  (higher  frequencies),  so  that  wave 
propagation  is  slowed  down  and  made  visible  to  the  human 
eye. 

5  NOTES ON OPERATION 

This  high-quality  instrument  fulfils  all  of  the  technical  re-
quirements  that  are  compiled  in  current  EC  guidelines.  The 
characteristics of this product qualify it for the CE mark. 
This instrument is only to be put into operation under special-
ist supervision in a controlled electromagnetic environment in 
research, educational and training facilities (schools, universi-
ties,  institutes  and  laboratories).  This  means  that  in  such  an 
environment,  no  mobile  phones  etc.  are  to  be  used  in  the 
immediate  vicinity.  The  individual  connecting  leads  are  each 
not to be longer than 2 m. 

Attention! 

Reviews: