8.0 CHECKING CALIBRATION

8.1

The BAT-12 thermometer can have a ‘worst case’ error of 0.2°C. The ‘worst

case’ is rarely met as it assumes that all sources of error are at their maximum, and
that all are additive.

A further possiblity of error is in the sensor. This is a maximum of 0.1°C at 50°C,
0.2°C at 100°C, rising to 3/8% of reading, i.e. a possible 1.5°C at 400°C. This is the
maximum error for Special Limits Copper-Constantan thermocouples permitted by
ANSI (American National Standards Institute.)

In practice, these errors are usually much lower unless poor measurement techniques
are being applied.

8.2

If errors are suspected, check F in Section 7 should first be made. Then the

instrument should be checked against a glass thermometer with 1/10th degree cali-
brations traceable to NIST.  Note that the calibration of a glass thermometer can be
invalidated by overheating  -- ensure that this never happens.

8.3 

Bind the thermocouple probe to the tip of the glass thermometer so that they

are close together. Do not compress the glass thermometer bulb.

8.4

Insert probe and thermometer in a Dewar flask or Thermos bottle containing

liquid at a suitable temperature. Do not allow the probe or thermometer to touch the
sides of the vessel. Ensure that the thermometer is immersed to its proper depth.
STIR WELL.

8.5

Allowing for some reading error in the glass thermometer (there can be none

in the digital BAT-12), readings should agree with the limits noted in 8.1. If readings
are outside these limits, the instrument should be returned for further checks. (See
section 10.0.) Recalibration of BAT-12 should not be attempted unless you have an
accurate temperature standard and a voltage standard such as a potentiometer with
one microvolt resolution.

A service manual is available. This includes circuit diagram, servicing information
and calibration instructions. Price is $25.

10

Summary of Contents for BAT-12

Page 1: ...ODUCTION 2 0 INITIAL INSTALLATION 3 0 OPERATING INSTRUCTIONS 4 0 BATTERY REPLACEMENT 5 0 ANALOG OUTPUT 6 0 TEMPERATURE MEASUREMENT WITH THERMOCOUPLE SENSORS 7 0 TROUBLE SHOOTING 8 0 CHECKING CALIBRATI...

Page 2: ......

Page 3: ...the same box if small or a separate package if they are longer than 12 inches 2 2 BAT 12 for battery operation is supplied ready for use with a 9V alkaline bat tery already installed Shipment will in...

Page 4: ...read temperature 4 0 BATTERY REPLACEMENT 4 1 When a colon appears in the center of the ditigal display the battery needs to be replaced However BAT 12 can be used without significant calibration error...

Page 5: ...he analog output with strip chart recorder The output impedance of BAT 12 is 2 Kilohms so connection to a potentio metric recorder will not affect either the calibration or accuracy The analog output...

Page 6: ...ronic reference circuitry 6 3 As compared with thermistor sensors which were formerly used exclusively in portable thermometers thermocouples have these advantages a Wide temperature range e g 200 C t...

Page 7: ...nical probes are guaranteed accurate to within 0 1oC in the range 0 50oC Copper Constantan type T thermocouples from other manufacturers are not normally this accurate Probes made from wire to special...

Page 8: ...1 2 handle suitable for use in the hand or in a micromanipulator and the micro thermocouple protrudes slightly beyond the steel sheath and has some thermal isolation from it This probe was designed f...

Page 9: ...og output Remove the output plug d Probe or sensor may be faulty open circuit Check by substituting anoth er probe or by a resistance check All probes should read less than 2000 ohms If you do not hav...

Page 10: ...ter can be invalidated by overheating ensure that this never happens 8 3 Bind the thermocouple probe to the tip of the glass thermometer so that they are close together Do not compress the glass therm...

Page 11: ...d Technology tables NIST monograph 125 Ambient Cold Junction 0 1 C from 0 to 50 C Temp Compensation Readout 3 1 2 digits 1 2 liquid crystal numerals Analog output 10mV per degree C approximately Power...

Page 12: ...itemp aol com 10 2 Physitemp Instruments Inc warrants this system to be free from defects in material or workmanship for 12 months from date of shipment Repair or replace ment will be made at no charg...

Page 13: ...PHYSITEMP INSTRUMENTS INC 154 Huron Avenue Clifton NJ 07013 U S A Tel 973 779 5577 Fax 973 779 5954 e mail physitemp aol com OPERATING MANUAL BAT 12 Microprobe Thermometer...

Reviews: