©2012 Physio-Control, Inc.
3
English
Basic Orientation
Chest Pad
Screen
Item Description
1
Power Button:
Press green button to turn device on. Press and hold for
2 seconds to turn device off.
2
Mute Button:
Press to turn metronome on or off.
In Event Review mode, press to switch between screens.
3
Palm Pad:
Hand placement area.
4
Airway Button:
Press to switch metronome between Airway
(continuous compressions) mode and No Airway mode.
5
Screen:
Displays compression feedback and other indicators.
6
Recoil Target Zone:
Changes from light green to dark green
when
compression is completely released.
7
Low Battery Indicator:
Appears when remaining battery capacity is
less than 30 minutes of operation.
8
Compression Rate Meter:
Displays rate as compressions per minute.
Changes to
Inactivity Timer
when compressions stop.
9
Elapsed Time Meter:
Displays elapsed time since beginning of event.
10
Airway Indicator:
Appears when device is in Airway mode.
11
Compression Target Zone:
Changes from light green to dark green
when compression reaches target depth of 5–6 cm (2–2.5 in).
12
Compression Depth History:
Dark green or gray segment in outer
circle shows depth of previous compression.
5
1
2
4
3
11
6
7
8
10
12
9