LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Boardstack
(continued)
Boardstack Installation
(continued)
12. Connect the 4-pin W10 Power/Therapy Cable connector to the A04 Therapy
PCB at J17.
13. Turn the device so that the ECG and therapy connectors are in view.
14. Connect the W25 Speaker Harness Extension Cable to the A01 System
PCB at J5.
15. Connect the ECG cable to the A02 PP PCB at J23.
16. Connect the W05 SpO2 Cable (if included) to the OEM PCB at J54, fasten
with a screw (255) using a 3/32 allen driver, and torque to 6.5 in-lb.
Note:
Carefully align the SpO2 and IrDA connectors to the sockets, and
gently press the connectors into the sockets using steady pressure to avoid
damage to the connector pins.
17. Connect the W03 IrDA Flex Cable to the A01 System PCB at J08, fasten
with a screw (255) using a 3/32 allen driver, and torque to 6.5 in-lb.
18. Install the grounding harness (219) from the power module to the EMI shield
(274) and support bracket (139) using new 4-40
0.312 screws (173);
torque to 6.8 in-lb.
19. Complete the process by
.
(Continued on next page)
9-78