Maintaining the Equipment
7-2
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
GENERAL MAINTENANCE AND TESTING
Periodic maintenance and testing of the LIFEPAK 20e defibrillator/monitor and accessories will
help detect and prevent possible electrical and mechanical discrepancies. If testing reveals a
possible discrepancy with the defibrillator or accessories, refer to
"General Troubleshooting Tips,"
. If the discrepancy cannot be corrected, immediately remove the defibrillator from
service and contact qualified service personnel. For testing information regarding accessories,
refer to the accessory operating instructions.
Each time you turn on the defibrillator/monitor, it performs self tests. If the defibrillator/monitor
detects a failure, the service LED illuminates.
A
MAINTENANCE DUE
message can be configured to appear on the screen at selected intervals
(3, 6 or 12 months) to remind you that the device is due for maintenance. The factory default is
OFF
, but it can be activated by service personnel.
Maintenance and Testing Schedule
lists the recommended maintenance and testing schedule. This schedule may be used
in conjunction with the internal quality assurance program of the hospital, clinic, or emergency
medical service where the defibrillator is used. An Operator’s Checklist is included in these
operating instructions (refer to Appendix D). Cables and paddles are a critical part of therapy
delivery and suffer wear and tear. Physio-Control recommends replacement of these
accessories every three years to reduce the possibility of failure during patient use.
Additional periodic preventive maintenance and testing, such as electrical safety tests,
performance inspection, and required calibration should be performed regularly by qualified
service personnel. The internal battery should be replaced by qualified service personnel every
two years.
Table 7-1
Recommended Maintenance Schedule
Operation
Daily
After
Use
As
Required
6
Months
12
Months
Daily auto test – performed automatically by defibrillator.
X
Complete Operator’s Checklist (refer to Appendix D).
X
Inspect defibrillator and CodeManagement Module.
X
X
Clean defibrillator and CodeManagement Module.
X
X
Check that all necessary supplies and accessories are
present (for example, gel, electrodes, ECG paper, etc.).
X
X
Perform user test.
X
Perform function checks:
Standard paddles monitoring check
X
Standard paddles defibrillation and synchronized
cardioversion check on battery power
X
Therapy cable monitoring check
X
Therapy cable defibrillation and synchronized
cardioversion check on battery power
X
Therapy cable pacing check
X
Preventive maintenance and testing.
X
Summary of Contents for LIFEPAK 20
Page 2: ...LIFEPAK 20e DEFIBRILLATOR MONITOR Operating Instructions ...
Page 3: ......
Page 4: ...LIFEPAK 20e DEFIBRILLATOR MONITOR OPERATING INSTRUCTIONS ...
Page 15: ......
Page 49: ......
Page 111: ......
Page 155: ......
Page 171: ......
Page 181: ......
Page 183: ......
Page 189: ......
Page 191: ......
Page 195: ......
Page 199: ......
Page 201: ......
Page 205: ......
Page 209: ......
Page 211: ......
Page 213: ......
Page 215: ......
Page 221: ......
Page 226: ......