LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Power/System PCB Cable (W01) Replacement
323
8
Power/System PCB Cable (W01) Replacement
Power/system PCB cable replacement consists of the following procedures:
•
Removing the Power/System PCB Cable (W01) (p. 323)
•
Installing the Power/System PCB Cable (W01) (p. 323)
Removing the Power/System PCB Cable (W01)
To remove the power/system PCB cable from the rear case (refer to
:
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Remove the system/therapy PCB assembly as described in
System (A01)/Therapy (A04) PCB
3.
Remove the OEM shield (
4.
For orientation, locate power PCB-J17 on the rear case diagram. (The system PCB J1 end of the
power/system PCB cable was disconnected as part of step 2.)
5.
Press the connector locking tabs, and then disconnect the power/system PCB cable from the
power PCB (W03) at J17.
Installing the Power/System PCB Cable (W01)
To install the power/system PCB cable into the rear case (refer to
Inside Rear Case Diagrams (p. 228)
:
1.
Connect the power/system PCB cable (
) to the power PCB (A03) at J17 (see
).
2.
3.
Install the system/therapy PCB assembly (described in
System (A01)/Therapy (A04) PCB Assembly Replacement (p. 232)
4.
Reassemble the case as described in
Reassembling the Case (p. 184)
.
P01
P17
Power PCB (A03)
System PCB (A01)
Figure 8.105—Power/system
PCB cable connections