LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Rear Case Replacement
316
8
Removing the Rear Case
To remove the rear case:
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Remove the system/therapy PCB assembly as described in
System (A01)/Therapy (A04) PCB Assembly Replacement (p. 232)
.
3.
Remove the screw (
) securing the clear, plastic, high-voltage shield (
from the interconnect bracket
4.
Remove the two mounting screws (
) from the capacitor bracket and remove the bracket (
) (see
). Discard the screws.
5.
Remove the two biphasic module mounting screws (
). Discard the screws.
6.
Observing orientation, cut the large tie wrap securing the A15 energy storage capacitor and interconnect bracket to the rear
case. Remove the energy storage capacitor, interconnect bracket, transfer relay and Biphasic module assembly
without
removing small cable ties and spade terminal connections
.
7.
Remove the OEM PCB as described in
Removing the OEM PCB (A06) (p. 262)
8.
Remove the parameter bezel as described in
Removing the Parameter Bezel (p. 311)
.
9.
Remove the NIBP/CO2 module as described in
Removing the NIBP (A21)/CO2 (A23) Modules (p. 288)
cover in this procedure is not required.)
10. Remove the SpO2 module as described in
Removing the SpO2 PCB (A16) (p. 275)
(if installed).
11. Remove the power PCB as described in
Removing the Power PCB (A03) (p. 254)
12. Remove the contact PCB as described in
Removing the Contact PCB (A07) (p. 359)
13. Remove the system connector cable and auxiliary connector cable as described in
Removing the System Connector Cable (W08)
or Auxiliary Power Cable (W09) (p. 332)
(use new O-ring seals (