LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
SpO2 PCB (A16) Replacement
275
8
Removing the SpO2 PCB (A16)
To remove the SpO2 PCB from the rear case (see
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Remove the system/therapy PCB assembly as
described in
System (A01)/Therapy (A04) PCB
.
3.
Remove the OEM shield (
4.
Disconnect the power/contact PCB cable (W05)
from the power board J12.
5.
Remove power/system cable (W01) from the power
board J17.
6.
Remove the power/therapy cable (W02) (
) from J8 on the power PCB.
J12
connection
J17
connection
J8
connection
Figure 8.67—Power/system PCB connections