LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
SpO2 PCB (A16) Replacement
279
8
12. To remove the SpO2 module from the bracket,
remove the four screws (
the SpO2 module to the mounting bracket. Discard
the screws.
13. Lift the SpO2 module off of the mounting bracket.
To remove the SpO2 PCB from the rear case (see
Figure 9.21 on p. 413
):
(Continued) 13 steps, (Page 5 of 5)
screw
locations
Figure 8.71—SpO2 PCB screw locations