LIFEPAK 15 Monitor/Defibrillator
Service Manual
Troubleshooting
User Interface Service Codes
122
5
User Interface Service Codes
Table 5.7—
Initial Digit 1, User Interface
Service
Codes (UI)
Code
Service Code Description
Corrective
Action Code
1005
UI_ERROR_DISPLAY_SELF_TEST (Self-test failed. Upper 16 bits of status code contain the expected CRC; lower 16 bits contain the
actual CRC.)
1006
UI_ERROR_ENERGY_FAULT (Defib charge out of 15% tolerance. Occurs only during manual mode.)
1007
UI_ERROR_12LEAD_KEY_SEEN (This unit is not configured to support 12-lead, but the software saw a key closure of this key.)
1008
UI_ERROR_ANALYZE_KEY_SEEN (This unit is not configured to support AED mode, but the software saw a key closure of this key.)
100a
UI_ERROR_NIBP_KEY_SEEN (This unit is not configured to support NIBP, but the software saw a key closure of this key.)
100b
UI_ERROR_CURRENTUP_KEY_SEEN (This unit is not configured to support pacing, but the software saw a key closure of this key.)
100c
UI_ERROR_CURRENTDOWN_KEY_SEEN (This unit is not configured to support pacing, but the software saw a key closure of this key.)
100d
UI_ERROR_RATEUP_KEY_SEEN (This unit is not configured to support pacing, but the software saw a key closure of this key.)
100e
UI_ERROR_RATEDOWN_KEY_SEEN (This unit is not configured to support pacing, but the software saw a key closure of this key.)
100f
UI_ERROR_PACER_KEY_SEEN (This unit is not configured to support pacing, but the software saw a key closure of this key.)
1010
UI_ERROR_PAUSE_KEY_SEEN (This unit is not configured to support pacing, but the software saw a key closure of this key.)
1037
UI_ERROR_UNCONFIGURED_BUTTON (Button test detected key that should not be present in current hardware configuration.)