LIFEPAK 15 Monitor/Defibrillator
Service Manual
Troubleshooting
Corrective Action Codes
141
5
11
Configuration or keypad Error:
a.
Check device configuration settings against installed hardware.
b.
Possible A09 Printer Control Keypad or A10 Main Keypad failure.
c.
Verify that the appropriate connecting cables and wire harnesses are functional.
8
Device manufacturing configuration bit is not set (Physio-Control service required).
Possible nonvolatile RAM failure on A01 System PCB.
13
Printer Error:
a.
Possible A12 Printer Assembly failure.
b.
Verify that the appropriate connecting cables and wire harnesses are functional.
15
Pacer out of calibration:
a.
Complete the TCP–Pacer Self-Calibration procedure, and then conduct Performance Inspection Procedures.
b.
Verify that the appropriate connecting cables and wire harnesses are functional.
c.
Possible A04 Therapy PCB failure.
17
Biphasic fault or unknown energy:
a.
Verify that the appropriate connecting cables and wire harnesses are functional.
b.
Possible A22 Biphasic PCB failure.
c.
Possible A04 Therapy PCB failure.
19
Dump error code (a02c):
a.
Verify that the appropriate connecting cables and wire harnesses are functional.
b.
Possible A04 Therapy PCB failure.
c.
Possible A22 Biphasic PCB failure.
d.
Possible A01 System PCB failure.
Table 5.17—Corrective Action Codes (Continued)
Corrective
Action Code
Description