LIFEPAK 15 Monitor/Defibrillator
Service Manual
Troubleshooting
Corrective Action Codes
140
5
Corrective Action Codes
Corrective action codes are referenced in the error code tables under
Service Log Code Categories (p. 119)
. If more than one action
is listed under Description, perform them in the order indicated.
Table 5.17—Corrective Action Codes
Corrective
Action Code
Description
1
System Communications or System Processing Error:
a.
Verify that the appropriate connecting cables and wire harnesses are functional.
b.
Possible EMI or ESD interference. Clear Service Log and then conduct Performance Inspection Procedures.
c.
Possible A01 System PCB failure.
2
Real time clock failure:
Possible nonvolatile RAM failure on coin battery power on A01 System PCB.
6
Therapy Processor Error:
a.
Verify that the appropriate connecting cables and wire harnesses are functional.
b.
Possible A04 Therapy PCB failure.
7
Power Processor Error:
a.
Verify that the appropriate connecting cables and wire harnesses are functional.
b.
Possible A03 Power PCB failure.
10
Defibrillator out of calibration:
a.
Complete the TCP–Defibrillator Energy Calibration procedure, and then conduct Performance Inspection Procedures.
b.
Verify that the appropriate connecting cables and wire harnesses are functional.
c.
Possible A04 Therapy PCB failure.
d.
Possible A22 Biphasic PCB/A14 Inductive Resistor failure.