Troubleshooting Tips
Table 23
Troubleshooting Tips for Defibrillation and Synchronized Cardioversion
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION
Charge time to 360 joules
exceeds 10 seconds
Battery low
•
Replace battery with fully
charged battery.
Operating temperature is too
low
•
Move patient and device to
warmer environment, if
necessary.
Energy not delivered to patient
when (shock) buttons are
pressed
Device is in Sync mode and
QRS complexes are not
detected
•
Adjust ECG size for optimum
sensing QRS or deactivate
SYNC
if rhythm VF/VT.
SYNC
accidentally pressed and
rhythm is VF/VT
•
Press
SYNC
to turn off Sync.
•
Press (shock) buttons.
Device in Sync
mode and
(shock) buttons not pressed and
held
until next detected QRS
•
Hold
(shock) buttons until
discharge occurs or next
detected QRS and
ENERGY
DELIVERED
message
appears
.
(shock) buttons pressed
before full charge reached
•
Wait for tone and message
indicating full charge.
Standard paddles connected
and (shock) button on
defibrillator front panel pressed
•
Simultaneously press
(shock) buttons on standard
paddles to discharge.
Sixty seconds elapsed before
(shock) buttons were pressed
after full charge. Energy was
internally removed.
•
Press (shock) buttons
within 60 seconds of full
charge.
Energy selection changed
•
Press
CHARGE
again.
CONNECT CABLE
message
appears
Therapy cable disconnected
during charging
•
Reconnect cable and press
CHARGE
again.
Therapy cable damaged
•
Replace therapy cable and
perform daily checks per
Operator’s Checklist.
ENERGY FAULT
message
appears (selected and available
energy)
Defibrillator out of calibration
•
Attempt to transfer energy.
•
Contact a qualified service
technician.
DISARMING
message appears
(shock) button not pressed
within 60 seconds after charge
complete
•
Recharge the defibrillator, if
desired.
Energy selected after charge
complete
•
Recharge the defibrillator.
SPEED DIAL
pressed
•
Recharge the defibrillator.
PACER
pressed
•
Recharge, if necessary, or
no action, if pacing desired.
Therapy electrodes or cable
disconnected
•
Reconnect electrode or
cable.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......