Available for Roger X (02) only
Pre-fitting
To ensure maximum benefit from the Roger system, program the sound processor with the recommended settings according to the table
below.
Model
Recommended sound processor setting
AB Naída CI
Set the Audio Mixing Ratio (Mic/Aux) to 50/50
AB Naída CI via ComPilot
Set the ComPilot Mixing Ratio to 50%
AB Neptune
Set the Audio Mixing Ratio (Mic/Aux) to 50/50
Cochlear Nucleus 7
Set Mixing Ratio to 1:1
Cochlear Nucleus 6 / 5
Set Mixing Ratio to 1:1 and select Autosensitivity
TM
+ ADRO configuration
MED-EL SONNET / Rondo / OPUS 2
None required
For
Cochlear’
s Nucleus 6, Nucleus 5 or the Advanced Bionics Neptune systems, Roger X* should be programmed. This will disable
AutoConnect and will ensure that the Roger X output impedances match the input impedance of the sound processor. A Roger
Touchscreen Mic or Roger inspiro is required for programming Roger X. Please follow the steps below to program the Roger X:
1.
Connect Roger X to the sound processor or ComPilot and switch it on.
Using Roger Touchscreen Mic:
2.
Turn Roger Touchscreen Mic on, hold it close to Roger X (less than 10 cm / 4 inches). Then go into
[Settings]
and select
[Roger
receiver settings].
3.
Select
[CI Profiles]
4.
Select the corresponding profile as shown in the table below.
5.
Select
[EasyGain level]
.
6.
Set the EasyGain to the corresponding value as shown in the table below.
Using Roger inspiro:
2.
Turn Roger inspiro on, hold it close to Roger X (less than 10 cm/4 inches) and select the function
[Check]
.
3.
Click
[Manage]
, scroll to
[CI module]
then click
[OK]
.
4.
Scroll to the corresponding setting as shown in the table below and click OK.
5.
Scroll to
[EasyGain]
and click
[OK]
.
6.
Set the EasyGain to the corresponding value as shown in the table below.
Model
Recommended CI Profile / CI Setting
for Roger X
Recommended EasyGain
for Roger X
AB Neptune with Roger X
Profile 4 / Setting 4
+8dB
Cochlear Nucleus 6 / 5 with Roger X
Profile 9 / Setting 9
0dB
MED-EL RONDO / OPUS 2 with Roger X
No programming required
Alternatively use Profile 2 / Setting 2
0dB
MED-EL SONNET with Roger X
No programming required
Alternatively use Profile 3 / Setting 3
0dB
No programming of Roger X is required for:
-
Advanced Bionics ComPilot
-
MED-EL audio processor
-
Cochlear Mini Microphone 2+
-
Older CI systems using MicroLink CI S
Roger 14, Roger 17, Roger 20 and Roger 21 are pre-configured for the compatible sound processor, that setting cannot be changed.