background image

Issue 

04/2020

S

y

s

t

e

m  I

n

s

t

r

u

c

t

i

o

n

s

®

IPROXX  SM/SM+/SMP

System Instructions

(Translation of the original 

System instructions)

04/2020

ID 56171 3.0

04/2020

ID 56171 3.0

®

IPROXX  SM/SM+/SMP

Summary of Contents for DewertOkin IPROXX SM

Page 1: ...Issue 04 2020 S y s t e m I n s t r u c t i o n s IPROXX SM SM SMP System Instructions Translation of the original System instructions 04 2020 ID 56171 3 0...

Page 2: ...OXX SMP 9 Function Diagram page IPROXX SMP 10 Function of the Display page 11 Maintenance and Repairs page 12 Trouble shooter s Guide page 13 Cleaning Disposal page 14 EC Declaration of Conformity pag...

Page 3: ...you are operating your system on battery power then observe the following notice Designated Use The Handset IPROXX is intended for connect with DEWERT drive control unit of Type CARE L HOSP L i e MCL...

Page 4: ...in control unit or a 1 DewertOkin drive with integral control PCB The keypad of the plugged in handset is completely locked Mode locked The Mode LED next to the lock doesn t glow the Power ON LED glow...

Page 5: ...vailable LED glows only when key is pressed with battery charging circuit LED flashes yellow when battery is charging LED glows constantly when battery is fully charged or is not connected during main...

Page 6: ...glow the Power ON LED glows dimly By turning the socket wrench to the symbol patient function the adjustment functions on the handset can be released these vary depending on the model Mode Patient The...

Page 7: ...er On Power On yellow Function Diagram IPROXX SM Socket wrench turned to handset is completely locked Mode LED doesn t glow Power ON LED glows dimly By turning the socket wrench to the adjustment func...

Page 8: ...able Battery powered operation mains voltage not available LED glows only when key is pressed with battery charging circuit LED flashes yellow when battery is charging LED glows constantly when batter...

Page 9: ...P with 10 adjustment keys Function LED Trendelenburg Antitrendelenburg Power ON LED Back of IPROXX SMP handset Rotary lock Socket wrench Back of handset Status LED Backrest up Footrest up down down St...

Page 10: ...he socket wrench to the symbol programming the adjustment functions on the handset can be programmed for the patient Mode Programming These functions are released or locked in pairs Programming y pres...

Page 11: ...ch turned to the programmed patient function adjustment functions on the IPROXX handset are released resp locked for the patient Status LED s of released functions glow yellow Status LED s of locked f...

Page 12: ...Status LED SMP LED s glow yellow the corresponding adjustment keys are released LED s do not glow the corresponding keys are locked Maintenance and Repairs At regular intervals carry out the inspecti...

Page 13: ...not connected The Function LED doesn t glow when a key is pressed control unit drive with first error security only Control unit drive or handset defective Contact your supplier dealer Problem Possibl...

Page 14: ...t allow fluids to penetrate the handset Damage to the handset could result Do not use a cleanser that contains benzene alcohol or similar solvents Disposal The IPROXX handset consists of electronic co...

Page 15: ...tic Compatibility Directive 2014 30 EU COMMISSION DELEGATED DIRECTIVE EU 2015 863 of 31 March 2015 amending Annex II to Directive 2011 65 EU of the European Parliament and of the Council as regards th...

Page 16: ...DewertOkin GmbH Weststra e 1 32278 Kirchlengern Germany Tel 49 0 5223 979 0 Fax 49 0 5223 75182 http www dewertokin de Info dewertokin de 04 2020 ID 56171 3 0 IPROXX SM SM SMP...

Reviews: