3
Handset locked
Handset released (SM/SM+/SMP)
released (SM+/SMP)
(SMP)
(SM/SM+/SMP)
Patient functions
Change Mode for nursing staff / care personnel
General symbols on the back of the handset
In order to guarantee single-fault protection you can lock the actuator movements via the
integrated locking device. In doing so, the motor circuits are disconnected by
electromechanic switch elements.
turn the locking switch to mode "locked",
check this by activating the adjustment key,
if locking is activated but movements will still be possible please exchange handset
for a new one
?
?
?
.
Information on battery power
?
If you are operating your system on battery power then observe the following notice.
Designated Use
The Handset
is
with DEWERT drive control unit of Type
CARE (L), HOSP (L), i.e. MCL, SGAG, etc.
IPROXX
intended for connect
®
The Handset
IPROXX
®
is
...
in the proximity
,
in an environment where
or
gases or vapours
(e.g. anaesthetics) are likely to occur,
in a damp environment, i.e. outdoors or
in beds intended for cleaning in wash tunnels.
not intended for use
high-frequency surgical equipment and defibrillators
inflammable
explosive
?
?
?
?
Note:
IPROXX
After it has been used to move your application (in the turnkey lock
positions or ), the Handset
handset should be
turned back to its locked position (turnkey lock position ). This
prevents the system from gradually discharging when the battery is
connected!
®
02/2013
ID No. 56171
02/2013
ID No. 56170