Ⅳ
. Instructions for use
•
Scope of application
:
Non-medical use. Hard surface of non-medical products, such as mobile phone, mouse of laptop, keyboard...,etc.
•
Precautions for Use
1. Take off all packaging material inside the chamber before use;
2. In the process of disinfection, If you need to open the door, please press the
“
stop
”
button before open;
3. The UV light cannot reach the contact part between the item's lower surface and the tray. If the bottom of the item needs to be
completely disinfected, please flip the item up and down once.
•
Requirements for object placement and disinfection plan
•
Solution to the lamp tube breakage
In case of accidental breakage of the lamp tube, it shall be ventilated for 30 minutes first. Please remove the debris with gloves,
put it in a sealed plastic bag and take it to a local waste disposal site for safe disposal.
Ⅴ
. Precautions
WARNING:
1. Keep a distance of at least 50 mm between the object and the lamp tube, and make sure every surface of the object is
exposed to ultraviolet light;
2. Objects shall not be stacked, and a gap of more than 60 mm shall be maintained between the objects;
3. Do not put wet objects or object with spilled liquids into the disinfection chamber;
4. The disinfection scheme of objects shall be based on the quantity and size of the objects and choose the corresponding
placing plan and disinfection time. See the below table for details.
•
Normal disinfection procedures are as follows
1. Make sure the power plug is plugged properly;
2. Open the chamber door, put the objects into the disinfection chamber, and place the objects according to the use manual;
3. Press
“
Power
”
button
4. Press the Timer to set the disinfection time according to above disinfection table;
5. Press the Start button for disinfection. Time for disinfection counts down as seconds;
6. During the disinfection process, the remaining time of the disinfection process can be seen through the timer. At the same ti me,
the specific situation in the disinfection chamber can be seen through the perspective window on the door to confirm that all the
lamps are working functionally (the perspective window can completely isolate ultraviolet radiation, and there is no need to worry
that ultraviolet radiation will penetrate the perspective window);
7. When the timer returns to zero, the whole disinfection process is finished;
8. Take out the objects and close the chamber door, and prepare for the next disinfection.
•
Do not operate the UV-C lamp when it is removed from the appliance enclosure.
•
This appliance contains a
UV-C lamp
.
•
Unintended use of the appliance or damage to the housing may result in the escape of dangerous UV-C
radiation. UV-C radiation may, even in little doses, cause harm to the eyes and skin.
•
Appliances that are obviously damaged must not be operated.
•
Read the maintenance instructions before opening the appliance.
•
If UVC products do not work, please contact the dealer for maintenance or replacement rather than
the customer.
Scenario
Min UV irradiance
4LOG disinfection
time
Precaution
Illustration
One small box (100mm*100mm*100mm)
0.96w/m2
(
96uw/cm2)
300sec
One big box (200mm*200mm*150mm)
0.5w/m2 (50uw/cm2)
600sec
3 small box
(
70mm*100mm*70mm
)
0.4w/m2 (40uw/cm2)
600sec
Object to object side keep
distance 50mm.
The object are at least 100mm
to left/right sides, 50mm to
front/back side.