. Technical Parameters
Ⅲ
. Product Installation
•
Preparation and inspection before installation
:
1. The installation position must be more than 30cm away from gas appliances, electric heating appliances or other heat sources,
or separated by board.
2. Before installation, the strength and surface flatness of the installation place must be checked, and the skew and dislocation of
the chamber door shall not appear.
3. It shall be installed in the place that can bear the weight of the chamber.
4. When installing, carefully check other factors that may cause the chamber to have malfunctions or accidents.
5. Impulse current is about 5 amperes, please check electrical capacity.
UVCC090 36W UVC Disinfection Chamber Instructions
Ⅱ
. Structure
Model
UVCC090 36W XX(01~20)
Dimensions
470*440*400mm
Volume
48L
Weight
9.4kg
Power
36W
Input Voltage
220-240V~
Frequency
50-60Hz
Fuse Protection
5A, 250V~
Disinfection Method
253.7 nm, UVC
Operating Temperature Range
10-40°C
Storage Temperature Range
-20-60°C
Ambient Humidity Range
10%-90%
UV Risk Group
The risk group of ultraviolet radiation in
the disinfection chamber is RG3, and the
outside of the chamber is RG0
•
WARNING
:
1.
Please ground reliably so as not to cause electrical shock accidents caused by poor grounding.
2. Make sure to disconnect the power supply before you move the disinfection chamber.
UVCC090