Press  and  “release”  the  AIR  button  and  press  the  vacuum  ON  button.    You
should  hear  the  vacuum  pull  the  door  sealed  shut.

When  the  vacuum  is  ready  (10  or  less  on  the  vacuum  gauge  and  RED  high
vacuum  light  is  off),  the  electron  beam  may  be  turned  on.

Shutting  down  the  SEM:

Turn  the  filament  current  to  zero.
Turn  off  the  high  tension  by  pushing  the  HT  button  and  letting  it  release.
Turn  the  black  detector  gain  down  1  turn  (clockwise).
Turn  off  the  SE  detector  (push  the  button  and  let  it  release).
Turn  the  spot  size  to  500  (this  is  the  lowest  lens  current  setting).
Turn  the  magnification  control  to  maximum  (lowest  lens  current  setting).
Turn  off  the  light  column  switch.
Turn  off  the  “marker”  switch.

The  SEM  can  be  left  in  this  setting  as  standby  mode  for  the  d a y .  

Do  not  leave  the  SEM  on  o v e r n i g ht .

To  completely 

shut  down  the  SEM  (overnight 

or  longer):

Press  the  vacuum  OFF  button.
Press  the  power  OFF  button.

Set  the  cooling  water  timer  to  at  least  30  minutes  to  cool  the  diffusion  pump
(quickly  turn  clockwise  past  OFF  to  >30).

If  the  microscope  will  not  be  used  again  for  an  extended  period  (1-2  days  or
more),  then  turn  off  the  master  mains  switch.

Please  be  sure  that  specimens  are  prepared  properly  and  are  free  of  volatile
substances.    Otherwise,  the  detector  and  final  pole  piece  aperture  will
become  contaminated  and  degrade  the  SEM  performance.

Summary of Contents for SEM 525

Page 1: ...E and adjust brightness and contrast for the display screens once set this does not change appreciably over time If using the Polaroid camera press and hold the photo start button and press the scan rate 2000 and 32 buttons This sets the scan rate for the camera screen The photo scan can be stopped by pressing the interrupt button Adjust operating parameters appropriate to the specimen characteris...

Page 2: ...isions the following should be expected at filament saturation Bias 1 25 kv saturated filament current 40 µA Bias 2 10 kv saturated filament current 35 µA To lower the saturated filament current increase the filament to Wehnelt distance Select spot size Generally speaking use a large spot size 200 or 500 for low magnification work under 1000 x magnification Smaller spot size is used for higher mag...

Page 3: ... 500 and minimum magnification turn up the filament current until the current no longer rises on the gauge Adjust the detector gain to provide a circular illumination Center this circle on the screen using the gun shift then use the gun tilt controls to give the brightest illumination Changing or inserting a new specimen Be sure the filament current and high tension is off and the filament has coo...

Page 4: ...magnification control to maximum lowest lens current setting Turn off the light column switch Turn off the marker switch The SEM can be left in this setting as standby mode for the day Do not leave the SEM on overnight To completely shut down the SEM overnight or longer Press the vacuum OFF button Press the power OFF button Set the cooling water timer to at least 30 minutes to cool the diffusion p...

Reviews: