background image

Use  gun  bias  1  for
20  and  greater  kv
and  bias  2  for  10  –
20  kv.

Use  a  spot  size  of  500
and  the  lowest
magnification  to  align
the  electron  gun.

Non-conductive  samples  may  charge  with  these  settings.    Biological
specimens  that  have  been  coated  usually  can  be  viewed  with  a  lower
accelerating  voltage,  usually  10  –  15  kV.    With  this  setting,  set  the  gun  bias
to  “2”  so  as  to  keep  the  emissions  current  in  a  safe  range  of  30  –  40 

µ

A.  A

lower  kV  setting  usually  improves  the  surface  image  but  also  adds  “noise”  to
the  detector  signal.

With  the  Wehnelt  gun  cap  fitted  with  a  200 

µ

  aperture,  and  the  filament  to

Wehnelt  distance  adjusted  6  small  divisions,  the  following  should  be
expected  at  filament  saturation:

Bias  1,  25  kv,  saturated  filament  current  =  40 

µ

A.

Bias  2,  10  kv,  saturated  filament  current  =  35 

µ

A.

To  lower  the  saturated  filament  current,  increase
the  filament  to  Wehnelt  distance.

Select  spot  size:

Generally  speaking,  use  a  large  spot  size  (200  or
500)  for  low  magnification  work  under  1000  x
magnification.    Smaller  spot  size  is  used  for  higher
magnifications.    As  spot  size  is  reduced,  the
detector  gain  must  be  increased  (and  visa  versa)
because  fewer  secondary  electrons  are  generated.
Black  level  is  not  affected  as  much.    For  5K  to  20K  magnification  work,  use  a
10  -  20 

µ

  spot  size.

Stage  controls:

Do  not  adjust  the  “Z”  control.    This  is  set  to  provide  an  optimal    distance
(10  –  12  mm  for  secondary  electron  detector  and  20  mm  for  the  EDAX
d e t e c t o r ) .

Do  not  force  any  stage  control  knob.

Summary of Contents for SEM 525

Page 1: ...E and adjust brightness and contrast for the display screens once set this does not change appreciably over time If using the Polaroid camera press and hold the photo start button and press the scan rate 2000 and 32 buttons This sets the scan rate for the camera screen The photo scan can be stopped by pressing the interrupt button Adjust operating parameters appropriate to the specimen characteris...

Page 2: ...isions the following should be expected at filament saturation Bias 1 25 kv saturated filament current 40 µA Bias 2 10 kv saturated filament current 35 µA To lower the saturated filament current increase the filament to Wehnelt distance Select spot size Generally speaking use a large spot size 200 or 500 for low magnification work under 1000 x magnification Smaller spot size is used for higher mag...

Page 3: ... 500 and minimum magnification turn up the filament current until the current no longer rises on the gauge Adjust the detector gain to provide a circular illumination Center this circle on the screen using the gun shift then use the gun tilt controls to give the brightest illumination Changing or inserting a new specimen Be sure the filament current and high tension is off and the filament has coo...

Page 4: ...magnification control to maximum lowest lens current setting Turn off the light column switch Turn off the marker switch The SEM can be left in this setting as standby mode for the day Do not leave the SEM on overnight To completely shut down the SEM overnight or longer Press the vacuum OFF button Press the power OFF button Set the cooling water timer to at least 30 minutes to cool the diffusion p...

Reviews: