background image

PageWriter TC20 Cardiographs

Wireless LAN Option

I

NSTALLATION

 I

NSTRUCTIONS

Edition 1

July 2011

Summary of Contents for PageWriter TC20

Page 1: ...PageWriter TC20 Cardiographs Wireless LAN Option INSTALLATION INSTRUCTIONS Edition 1 July 2011 ...

Page 2: ...ntained in it is strictly reserved for current Philips personnel and Philips customers who have a current and valid license from Philips for use by the customer s designated in house service employee on equipment locat ed at the customer s designated site Use of this document by unauthorized persons is strictly prohibited Report violation of these requirements to the Philips Legal Department This ...

Page 3: ...ons and security settings Additional information about the AmbiCom adapter is available at the AmbiCom website www ambicom com This document provides the following information Security Options 2 Configuring Wireless Adapter Settings 4 Before You Begin 4 Step 1 Inserting the Wireless USB Adapter 5 Step 2 Loading Authentication Certificates and Keys 6 Loading Authentication Certificates 7 Importing ...

Page 4: ...n Table 1 You configure all wireless LAN security and connection options on the network configuration screens accessed through the Setup button on the main tool bar NOTE Some authentication methods may require that a valid authentication certificate be loaded and enabled on the cardiograph For details see Step 2 Loading Authentication Certificates and Keys on page 6 Table 1 Popular EAP types and r...

Page 5: ...nd how the encryption key will be created None No encryption WEP Wireless Equivalency Privacy WEP with up to four static keys that can be 40 bit 128 bit in ASCII or 128 bit in hex WPA PSK TKIP with PSK either ASCII passphrase PSK or hex PSK WPA TKIP or AES TKIP or AES with key generated during EAP authentication WPA2 PSK AES with PSK either ASCII passphrase PSK or hex PSK WPA2 TKIP or AES TKIP or ...

Page 6: ... are described in Table 1 Popular EAP types and required credentials on page 2 Contact your network administrator for any certificate files The PageWriter TC Cardiograph Wireless LAN Upgrade Kit includes the AmbiCom adapter and information about the adapter Configuration Step See 1 Install the wireless adapter If your cardiograph already has the adapter installed proceed to step 2 Step 1 Inserting...

Page 7: ...nectivity To enable the cardiograph for a wireless connection Insert the AmbiCom USB adapter USB dongle into the USB port on the side of the cardiograph If the port is already in use that is a USB memory stick with ECGs or worklist entries is currently in the port remove the memory stick and replace it with the wireless USB dongle The dongle illuminates when it is properly connected You can now lo...

Page 8: ...ficate and private key located in the NetUI utility s certificate directory into the Microsoft System store Table 3 Load Certs and Import Certs options If the encryption method you are using requires a certificate or an optional private key have the network administrator generate the certificate private key as required and load these files onto a USB memory stick CAUTIONS Do not insert a USB memor...

Page 9: ...n the main screen touch Setup on the tool bar The Configuration Setup and Service Utilities menu appears 5 Touch Configure ECG Network Settings 6 Touch the Wireless Network tab The Wireless LAN Adapter Settings section provides access to the NetUI configuration utility and buttons for loading certificates 7 Touch Load Certs NOTE If the USB stick is not inserted in the USB port an alert dialog box ...

Page 10: ...tificates and Private Keys next Importing User Certificates and Private Keys NOTE This step is only required when using EAP TLS authentication To import user certificates and private keys 1 On the main screen touch Setup on the tool bar 2 Touch Configure ECG Network Settings then touch the Wireless Network tab 3 In the Wireless LAN Adapter Settings section touch Import Certs 4 In the Certificates ...

Page 11: ...icate File browser navigate to the Certificates folder by selecting in order Storage Card Philips Certificates touching OK after each selection Then select the certificate file and touch OK The next step is to import the private key associated with the certificate 8 To import the key do the following a In the Certificates window select the My Certificates directory again The certificate you just i...

Page 12: ... After installing the wireless card and loading certificates if needed as described in the previous sections you must select the SSID for the network connection to use choose encryption authentication methods to use and specify any required credentials to permit access NOTE If you are going to use EAP TLS authentication proceed directly to Configuring a WPA or WPA2 EAP TLS Connection on page 15 Se...

Page 13: ...ot show the entry you need it is either because the connection is not on the network or perhaps the SSID is not being broadcast Regardless of the cause you must add the connection to configure it To add a connection 1 Ensure the AmbiCom wireless USB dongle is inserted in the USB port on the cardiograph 2 On the main screen touch Setup on the tool bar 3 Touch Configure ECG Network Settings then tou...

Page 14: ...selecting the SSID as described on page 10 select the following options in the Wireless Network Properties dialog box 1 Touch the Encryption dropdown arrow and select the desired method AES or TKIP 2 Touch the Authentication dropdown arrow and select WPA PSK or WPA2 PSK 3 Touch the Network key field and type the key 4 Touch OK The Wireless Information tab Status field refreshes and should show tha...

Page 15: ... or WPA2 PEAP MSCHAPv2 Connection This option uses a single certificate on the cardiograph for authentication To configure a WPA or WPA2 PEAP MSCHAP connection After selecting the SSID as described on page 10 select the following options in the Wireless Network Properties dialog box 1 Touch the Encryption dropdown arrow and select the desired method AES or TKIP 2 Touch the Authentication dropdown ...

Page 16: ...K in the top right corner of the window The Wireless Information tab Status field refreshes and should show that the cardiograph is connected to the selected network and give an indication of signal strength 8 To specify multiple networks to which the cardiograph can connect repeat step 5 on page 11 and the steps in this procedure as often as needed to configure the necessary set of networks Confi...

Page 17: ... required certificate and key page 7 2 Ensure the AmbiCom wireless USB dongle is inserted in the USB port on the cardiograph 3 In the Wireless LAN Adapter Settings section touch Configure The NetUI utility appears displaying available networks in the Wireless Information tab The list refreshes automatically 4 Select the SSID of the desired wireless network and touch Connect Scroll the list if need...

Page 18: ...ertificate and touch OK The Authentication Settings dialog box appears again showing details about the selected certificate 11 Touch the Validate Server check box to clear it and touch OK 12 In the Wireless Network Properties dialog box touch OK 13 When prompted to log on provide the certificate user name and domain 14 Touch OK in the top right corner of the window ...

Page 19: ...on is complete Proceed to Step 4 Restarting the Cardiograph on page 18 Configuring a WEP Connection This option uses a network key and a key index To configure a WEP connection After selecting the SSID as described on page 10 select the following options in the Wireless Network Properties dialog box 1 Touch the Encryption dropdown arrow and select the desired method WEP 2 Touch the Authentication ...

Page 20: ...t step 5 on page 11 and the steps in this procedure as often as needed to configure the necessary set of networks Configuration is complete Proceed to Step 4 Restarting the Cardiograph on page 18 Step 4 Restarting the Cardiograph You must restart the cardiograph for the new wireless settings to be applied To restart the cardiograph 1 Press and hold down the On Standby button for five seconds to sh...

Page 21: ...tus Indicator on Status Bar Wireless LAN Connectivity Indicator Table 5 Wireless LAN Connectivity Indicator Label Description Wireless LAN Connectivity Wireless LAN connectivity Indicates the signal strength of a the wireless LAN connection Green bars indicate that the wireless adapter is associated to an access point The higher the number of green bars the better the connection Gray bars indicate...

Page 22: ...eless dongle Dimensions 1 39 x 0 62 x 0 31 in 35 5 x 16 x 8 mm System Interface USB port 1 0 and 2 0 Antenna 2dBi chip antenna One transmit one receiving 1T1R Output Power 1 dBm 11n g 14 dBm 11b 17 dBm Operating Temperature 0 C to 40 C 32 F to 104 F Humidity 90 max noncondensing Network Standards 802 11g up to 54 Mbps 802 11b up to 11 Mbps Regulatory Domain Support FCC Americas parts of Asia and M...

Page 23: ...ent Privacy WEP Wi Fi Protected Access WPA WPA2 802 1x authentication Encryption Wireless Equivalent Privacy WEP RC4 Algorithm Temporal Key Integrity Protocol TKIP RC4 Algorithm Advanced Encryption Standard AES Rijndael Algorithm Encryption Key Provisioning Static 40 bit and 128 bit lengths Pre Shared PSK Dynamic 802 1X Extensible Authentication Protocol Types EAP TLS PEAP MSCHAPv2 ...

Page 24: ... 300 328 EN 301 489 EN 60590 EN 50371 EU 2002 95 EC RoHS FCC Regulatory Domain FCC Subpart B Class B FCC Subpart C Part 15 247 15 207 ANSI C63 4 2003 Industry Canada RSS 210 TELEC Regulatory Domain RCR STD 33 ARIB STD T66 ARIB STD T71 Certifications Wi Fi Alliance 802 11b 802 11g WPA Enterprise WPA2 Enterprise Cisco Compatible Extensions CCX Version 4 ...

Page 25: ...on the Philips Response Center go to www medical philips com main services response_center North America Response Centers Country Telephone Number Canada 800 323 2280 Mexico 01 800 710 8128 Puerto Rico 1 787 754 6811 United States 800 722 9377 South America Response Centers Country Telephone Number Argentina 54 11 4546 7698 Brazil 0800 701 7789 Chile 0800 22 3003 Columbia 01 8000 11 10 10 Peru 51 ...

Page 26: ...5 Finland 358 615 80 400 France 0 810 835 624 Germany 0180 5 47 5000 Greece MCR Response Center located in The Netherlands 31 40 2781619 Hungary MCR Response Center located in The Netherlands 31 40 2781619 Italy 0800 232100 Netherlands 31 40 27 211 27 Norway 47 800 84 080 Poland MCR Response Center located in The Netherlands 31 40 2781619 Rumania MCR Response Center located in The Netherlands 31 4...

Page 27: ...0 4050 Sweden 46 200 81 00 10 Switzerland 0800 80 3000 German 0800 80 3001 French Asia Response Centers Country Telephone Number Australia 1800 251 400 China 800 810 0038 Hong Kong 852 2876 7578 India 1600 112 444 Indonesia 62 21 7910040 ext 8610 Japan 81 0120 095 205 Korea 82 02 3445 9010 Malaysia 1800 886 188 New Zealand 0800 251 400 Philippines 63 2 8162617 ext 875 Europe Response Centers conti...

Page 28: ... Installation Instructions Singapore 1800 Philips Taiwan 0800 005 616 Thailand 02 614 3569 Africa and Middle East Country Telephone Number All countries MCR Response Center located in The Netherlands 31 40 2781619 Asia Response Centers continued Country Telephone Number ...

Page 29: ......

Page 30: ...an Road Andover MA 01810 USA Copyright 2011 Koninklijke Philips Electronics N V All rights reserved PageWriter TC20 Cardiograph Wireless LAN Installation Instructions Printed in USA 453564267741 Edition 1 July 2011 453564267741 Edition 1 ...