Measurement Specifications
12-16
Safety Standards & Specifications
Measurement Specifications
ECG
Parameter
Specification
ECG channel
transmitted Leads
•
3 electrodes
•
5 electrodes
•
5 electrodes,
EASI
•
6 electrodes
•
Channel #1 = I, II, or III
•
Channel #1 = II
Channel #2 = III
Channel #3 = MCL
•
Channel #1 = Va-i
Channel #2 = Va-s
Channel #3 = Ve-s
•
Channel #1= II
Channel #2 = III
Channel #3 = Va
Channel #4 = Vb
Resolution
5
μ
V
ECG Input
Differential, defibrillator protected against 360
joules discharge into a 100 ohm load
Input Impedance
> 5 megohms (@ 10 Hz
Input Dynamic Range
+/- 9 mV
DC Offset Range
+/- 320 mV
CMRR
> 90 dB @ 50, 60 Hz
Ban/- 3 dB
0.05 to 40 Hz