HEARTSTART M5066A
Patient treatment protocol options
parameter
settings
default
default description
“call EMS” voice
reminder timing
• At power on (when
the user turns on
the HeartStart)
• At power on and at
the start of the first
patient care pause
• At the start of the
first patient care
pause
• No reminder
At the start of
the first
patient care
pause
Provides a voice reminder to make
sure emergency medical services have
been called, at the start of the first
patient care pause.
shock series
1, 2, 3, 4
1
The automatic protocol pause for CPR
is activated each time a shock is
delivered.
*
During the protocol pause, the
HeartStart does not perform rhythm
analysis.
The length of the protocol pause after
a shock series is completed is
determined by the protocol pause
timer setting.
shock series interval
(minutes)
1.0, 2.0,
∞
(infinity)
1.0
A delivered shock must occur within 1
minute of the previous shock to be
counted as part of the current shock
series.
NOTE: This parameter is only applicable
when the shock series is not configured to
the default 1 shock.
* A shock series begins when a shock is delivered after the HeartStart is turned on. A new shock series begins after a
protocol pause. If shock series is configured for 2 or more, a new shock series also begins if the time since the previous
shock exceeds the shock series interval setting.
Summary of Contents for HEARTSTART
Page 2: ...Philips Medical Systems Intentionally blank ...
Page 4: ...Philips Medical Systems Intentionally blank ...
Page 5: ...Philips Medical Systems HeartStart Defibrillator M5066A QUICK REFERENCE ...
Page 6: ...Philips Medical Systems Intentionally blank ...
Page 8: ...Philips Medical Systems Intentionally blank ...
Page 10: ...Philips Medical Systems Intentionally blank ...
Page 26: ...Philips Medical Systems HEARTSTART M5066A Notes ...
Page 56: ...Philips Medical Systems HEARTSTART M5066A Notes ...
Page 65: ...Philips Medical Systems Intentionally blank ...
Page 66: ...Intentionally blank Philips Medical Systems ...
Page 67: ...Philips Medical Systems Intentionally blank ...
Page 68: ...Philips Medical Systems Intentionally blank ...