Technical information
HeartStart Defibrillator specifications
The specifications provided in the following tables are nominal values. Additional
information can be found in the
Technical Reference Manual
for HeartStart
Automated External Defibrillators, located online at www.medical.philips.com.
Physical
Environmental
category
specifications
size
2.80” H x 7.40” D x 8.30” W (7.1cm H x 19cm D x 21cm W).
weight
Approximately 3.3 lbs (1.5 kg) with battery and pads cartridge installed.
category
specifications
temperature and
relative humidity
Operating (battery and pads cartridge installed):
32° to 122° F (0° to 50° C)
0% to 95% RH (non-condensing).
Standby (battery and pads cartridge installed):
50° to 109° F (10° to 43° C)
10% to 75% RH (non-condensing).
Storage/shipping (with battery and pads cartridge):
-4° to 140° F (-20° to 60° C) for up to 2 days
0% to 85% RH (non-condensing)
altitude
Operates at 0 to 15,000 feet; can be stored at up to 8,500 feet, in standby
mode.
shock/drop abuse
tolerance
Withstands 1 meter drop to any edge, corner, or surface.
vibration
Operating: meets EN1789 random, road ambulance.
Standby: meets EN1789 swept sine, road ambulance.
Summary of Contents for HEARTSTART
Page 2: ...Philips Medical Systems Intentionally blank ...
Page 4: ...Philips Medical Systems Intentionally blank ...
Page 5: ...Philips Medical Systems HeartStart Defibrillator M5066A QUICK REFERENCE ...
Page 6: ...Philips Medical Systems Intentionally blank ...
Page 8: ...Philips Medical Systems Intentionally blank ...
Page 10: ...Philips Medical Systems Intentionally blank ...
Page 26: ...Philips Medical Systems HEARTSTART M5066A Notes ...
Page 56: ...Philips Medical Systems HEARTSTART M5066A Notes ...
Page 65: ...Philips Medical Systems Intentionally blank ...
Page 66: ...Intentionally blank Philips Medical Systems ...
Page 67: ...Philips Medical Systems Intentionally blank ...
Page 68: ...Philips Medical Systems Intentionally blank ...