Safety Considerations
M3500B HeartStart XLT Defibrillator/Monitor
13-21
13
WARNING
Avoid contact between the patient and conductive fluids and/or metal objects, such
as the gurney. Contact with metal objects could cause unintentional current path-
ways.
WARNING
Operating the HeartStart XLT or its accessories in conditions outside the environ-
mental specifications can result in device or accessory malfunction.
WARNING
Avoid exposing the display to direct sunlight on a hot day. Overheating can occur and
cause the display to black out, making the HeartStart XLT temporarily unusable.
WARNING
Medical electrical equipment which does not incorporate defibrillator protection
should be disconnected during defibrillation.
WARNING
Electric shock hazards exist internally. Do not remove assembly screws. Refer ser-
vicing to qualified personnel.
CAUTION
This device has not been evaluated for use with electrosurgery equipment.