
Service Modes, Error Codes and Fault Finding
EN 34
5.
(first connect ground to measuring equipment, than measure
the gate).
5.8.6
Standby Supply
The simplest way is to replace the components of the Standby
Supply with repair kit 3122 785 90460.
5.8.7
Line Deflection
The simplest way is to replace the components of the Line
Deflection circuitry with repair kit 3122 785 90330.
5.8.8
Frame Deflection
Caution: When the Frame Deflection circuitry is suspected,
one must be careful. Because there is a DC-voltage on the
frame deflection, the beam current could damage the CRT
neck, leading to a defective CRT.
•
Best thing to do is:
•
Interrupt the resistors 3403 and 3404 on the CRT panel
(diagram F1), in order to remove the ‘filament’ voltage from
the tube (no beam current, so no chance of destroying the
CRT).
•
Interrupt resistor 3403 on the LSP (diagram A4) to disable
the ‘SUP-ENABLE’ line.
•
Measure with a multi-meter, or better with a oscilloscope,
the functionality of the Frame stage.
•
After you have found the cause, exchange the defective
component (e.g. TDA8177), and re-solder the interrupted
resistors.
Table 5-4 Repair tips
Phenomenon
Possible Cause
Repair tip
No picture, no LED.
Standby Supply defective.
Measure circuitry (see diagram A2). Start at test-point A19.When the Mains switch
is 'on', this voltage must always be available
No picture, no LED.
Bad connection of SSB in
SIMM-connector.
In case of a bad connection, it can happen that there is no picture, and that the
Standby LED is not controllable. Re-place the SSB.
No picture, red LED
blinking at 3 Hz.
Set is in protection due to
various causes. For error
codes see error-code list.
You have no picture, so:
Read the error buffer via ComPair (error buffer is accessible when set is in protec-
tion, compare-file will guide you to this)
Read the blinking LED information via standard remote command
<mute>06250X<ok>
Or you read the error code sequence via standard remote command
<mute>062500<ok>. When you have found the error, check the circuitry related to
the supply voltage and I
2
C-communication or the circuitry that triggers the protec-
tion.
No picture, red LED
blinking with code 6-6-6
or 18-18-18.
No communication on slow
I
2
C- or fast I
2
C-bus.
As processor cannot communicate with one of the 2 busses it the standby-led spon-
taneously starts blinking 6-6-6 . . . or 18-18-18 . . .. If in the error buffer somewhere
is an error 6 or 18, these will have the highest priority starting the mentioned blink-
ing. Measure dependent of the error on the I
2
C-bus which device is loading the bus.
(Use I
2
C -overview)
No picture, red LED
blinking with code 1-1-1.
No communication on NVM-
I
2
C bus to the µP.
As the µP cannot communicate with the NVM I
2
C bus, it spontaneously starts blink-
ing 1-1-1. Note: when there is no access to the NVM, a lot of picture setting can go
wrong.
No picture, no sound.
Set is making audible
squeaking sound
Supply is possibly in hiccup-
mode, which is audible via a
squeaking supply transformer.
Possible causes:
V
BAT
is shorted (caused by short circuited line transistor 7421), or
Sound winding is shorted (amplifier is shortening the power supply lines), or
D6514 is shorted (due to a too high V
BAT
).
Remove excessive load, to see what causes the failure, or check feedback circuit.
See repair tip 'Main Power Supply' (supply needs a minimal load).
No picture, no sound.
Green LED works fine
Supply does not work correct-
ly.
If e.g. V
BAT
is only about 90 V, it is possible that the regulator IC (7506) is defective.
No RC-reception. Red
LED does not echo RC-
commands.
µP circuitry or RC-receiver is
defective.
In case the set does react on a local keyboard operation, you must check the RC-
receiver circuitry (diagram E).
Relay (degaussing) is
not audible, when set is
switched from 'off' or
'standby' to 'on'.
µP is not working correct-
ly.When pin 115 is low, the de-
gaussing must be activated.
Check RESET-circuitry on diagram B5. Check the level on pin 115 when you switch
the set 'on'. Signal must be low initially and go to high after approx. 12
µ
s.
No sound, only picture. Measure A7/A8 on diagram
A1. Possibly the audio amplifi-
er is defect (but not shorted),
or the 'sound enable' line is
high (diagram A5). Further,
measure the audio signal path
(HIP-MSP-Switch ICs-Amplifi-
er).
In ComPair, a beep test can determine where the signal stops (use loudspeakers
or headphone).Possible cause is also the 'PROT1' protection (see diagram. A6).
This is a loudspeaker DC protection, which the µP cannot sense. Via circuitry 7701/
7707, this signal can put audio amplifiers 7700 and 7702 in standby mode (tri-state
input). Warning: Find out, which amplifier is causing the problem, because it is pos-
sible, that the other one is OK.
Summary of Contents for EM5E
Page 7: ...Directions for Use EN 7 EM5E 3 3 Directions for Use ...
Page 8: ...Directions for Use EN 8 EM5E 3 ...
Page 9: ...Directions for Use EN 9 EM5E 3 ...
Page 10: ...Directions for Use EN 10 EM5E 3 ...
Page 11: ...Directions for Use EN 11 EM5E 3 ...
Page 12: ...Directions for Use EN 12 EM5E 3 ...
Page 13: ...Directions for Use EN 13 EM5E 3 ...
Page 14: ...Directions for Use EN 14 EM5E 3 ...
Page 15: ...Directions for Use EN 15 EM5E 3 ...
Page 16: ...Directions for Use EN 16 EM5E 3 ...
Page 17: ...Directions for Use EN 17 EM5E 3 ...
Page 18: ...Directions for Use EN 18 EM5E 3 ...
Page 19: ...Directions for Use EN 19 EM5E 3 ...
Page 20: ...Directions for Use EN 20 EM5E 3 ...
Page 36: ...Service Modes Error Codes and Fault Finding EN 36 EM5E 5 Personal Notes ...