Menu: Settings and alarms
Mode
CPAP, S, S/T, PC, T
Flex Lock (only in S mode)
On (patient cannot adjust Flex level)
Off (patient can adjust Flex level)
Flex (only in S mode)
Off/1/2/3
AVAPS
On/off
AVAPS rate
From 0.5 cmH
2
O/min up to
5 cmH
2
O/min per 0.5 cmH
2
O/min step
Tidal volume
(only when AVAPS is enabled)
200 - 1500 ml
per 10 ml step
IPAP max
(only when AVAPS is enabled)
IPAPmin - 30 cmH
2
O
per 0.5 cmH
2
O step
IPAP min
(only when AVAPS is enabled)
4 - IPAPmax
per 0.5 cmH
2
O step
EPAP
4 - 25 cmH
2
O per 0.5 cmH
2
O step
Breath rate
0 to 40 bpm (4-40 bpm for T mode)
Inspiratory time
0.5 to 3 sec. per 0.1 sec. step
Rise time lock
On (patient cannot adjust rise time)
/Off (patient can adjust rise time)
Rise time
1 (fastest) to 6 (slowest)
Ramp length
0 to 45 minutes per 5 minutes step
Ramp start pressure
From 4 cmH
2
O to prescribed EPAP per
1 cmH
2
O step
System One humidification
On (System One humidity control)/
Off (classic mode)
Humidifier
1 (coolest) to 5 (hottest)
Humidifier
Ramp
cm
H
2
O
cm
H
2
O
ml
l/min
l/min
BPM
S/T AVAPS: 0.5
0
5
10
15
20
25
30
Menu
Pressure
12.1
Vte
325
Leak
30.0
MinVent
3.9
RR
12
I:E Ratio
1:3.0
A
1
AVAPS activated
with a speed rate
of 0.5 cmH
2
O/min.
Breath
indicator
(patient or
machine settable
in Options)
Device
unlocked
Humidifier
connected,
setting 1
Pressure bar graph
Real-time
monitored values
when detailed
view ON (Options
settings)
Mode of ventilation
External battery
connected and
in use
SD Card
inserted
Symbols and icons
FLEX enabled
FLEX enabled and FLEX setting
(eg. 1)
Humidifier connected and
Humidifier setting (eg. 1)
Bad humidifier state
(flashing symbol displays)
Ramp activated
SD Card error
(bad memory card inserted)
AVAPS
AVAPS: 1
FLEX
FLEX: 1
A
1
AVAPS
AVAPS: 1
FLEX
FLEX: 1
A
1
AVAPS
AVAPS: 1
FLEX
FLEX: 1
A
1
AVAPS
AVAPS: 1
FLEX
FLEX: 1
A
1
AVAPS
AVAPS: 1
FLEX
FLEX: 1
A
1
AVAPS
AVAPS: 1
FLEX
FLEX: 1
A
1
AVAPS
AVAPS: 1
FLEX
FLEX: 1
A
1
AVAPS
AVAPS: 1
FLEX
FLEX: 1
A
1
AVAPS
AVAPS: 1
FLEX
FLEX: 1
A
1
AVAPS
AVAPS: 1
FLEX
FLEX: 1
A
1
AVAPS
AVAPS: 1
FLEX
FLEX: 1
A
1
AVAPS
AVAPS: 1
FLEX
FLEX: 1
A
1
AVAPS
AVAPS: 1
FLEX
FLEX: 1
A
1
AVAPS
AVAPS: 1
FLEX
FLEX: 1
A
1
Apnea alarm enabled
Audio pause is active
Circuit disconnect
alarm enabled
External battery is 80% capacity
External battery is 20% capacity
External battery has less than
20 minutes left
External battery has less than
10 minutes left
External battery is empty
Getting started
1
Display screen
Perform action specified on screen
(humidifier setting, cancel action
or exit menu)
Navigate menus or change settings
Perform action specified on screen
(activate ramp, select/modify/
confirm parameters)
Audio pause button
Start/stop button
6
5
4
3
2
1
Humidifier
Ramp
cm
H
2
O
S/T AVAPS: 0.5
0
5
10
15
20
25
30
Menu
12/30/2010
06:12 PM
A
1
1
2
3
4
5
6
How to stop
How to start therapy
Device navigation
• Connect the
device to AC
or DC power
source
• Press to
switch on the
device
• Select
to start
the treatment
Navigate
OK
Cancel
cm
H
2
O
cm
H2O
ml
l/min
S/T
0
5
10
15
20
25
30
Pressure
3.5
Vte
0
ml
RR
20
Leak
0.0
Standby
Power Off
S/T AVAPS: 0.5
Menu
Setup
Therapy
A
12/30/2010
06:12 PM
• Press
• To keep power on but stop
the blower, select ‘Standby’
• Then select
• To turn power off, select
‘Power Off’
• Then select
How to unlock
To unlock the device
• Press and hold and
for a few seconds to access
Setup page
• Scroll down and select ‘Options’
• Modify ‘Menu access’ from
‘Limited’ to ‘Full’
• Press
to validate
• Press
and
to come back to
the main screen
When
appears on the main
screen, the device is unlocked.
cm
H
2
O
cm
H2O
ml
l/min
l/min
BPM
S/T AVAPS: 5.0
0
5
10
15
20
25
30
Menu
Pressure
12.6
Vte
0
Leak
0.0
MinVent
0.0
RR
11
I:E Ratio
1:1.0
To navigate through the Options
menu:
• Use
to scroll
through the menu.
• Use the left and right
buttons to perform the actions
specified on the on-screen buttons
and set according to your preference.
Menu: Options
Menu: Alarm and Event Logs
Menu: Settings and alarms
(cont.)
Tubing type lock
On (patient cannot adjust tubing size
diameter)/Off (patient can adjust
tubing size diameter)
Tubing type
15 mm/22 mm
System One resistance lock
On (patient cannot set mask type)/
Off (patient can set mask type)
System One resistance (mask type)
0, X1, X2, X3, X4 or X5
Circuit disconnect alarm
Off; 15 or 60 sec.
Apnea alarm
Off; 10, 20 or 30 sec.
Low tidal volume alarm
(only when AVAPS is enabled)
On/off
Low minute ventilation alarm
Off; 1 to 99 lpm
High respiratory rate alarm
Off; 4 to 60 BPM
Menu access
Full (device unlocked)/Limited (device locked)
Detailed view
On (monitored data appear on the main screen)/Off
(no monitored data on screen, but time and date)
Language
You have the choice between 22 different languages
Pressure units
cmH
2
O/hPa
Breath indicator
Patient (spontaneous breath)
/machine (controlled breath)
Keypad lock
On (buttons are inactive)/off (buttons are active)
Keypad backlight
On/off
LCD brightness
1 (darkest) to 10 (lightest)
Screen saver
Dim (backlight is decreased)/Breath (black screen,
but manometer and breath indicator are visible)/
Off (no screen saver)
Date & time format Date and time settings
Blower hours
Number of hours the blower has been active
(resettable)
Therapy hours
Number of hours the patient receives therapy
(resettable)
Menu access
Full
Detailed view
ON
Language
English
Pressure units
cmH
2
0
Breath Indicator
Patient
Menu
Options
S/T AVAPS: 0.6
1/18
Navigate
Setup
Modify
Finish
A
Under ‘Options’, you have access to the following parameters:
Alarm Log
The alarm log lists the 20 most recent alarms or messages that
appeared on the device display.
Event Log
The event log displays a list of all events that have occurred,
in chronological order.