Technical information
EMC information
The Philips Avent breast pump needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided in this section.
Portable and mobile RF communications equipment can affect the Philips Avent breast pump.
The electronic breast pump has no essential performance. Due to EMC influences, like wireless home
network devices, mobile phones or cordless telephones, the breast pump can turn off or go into error
mode. This will not lead to unacceptable risks. To prevent interference, keep other electric equipment
at least 3.3 feet/1.0 m away from the breast pump during expressing and do not stack with other
electrical equipment. Cable length of the adapter: 2.50 meter/ 8.2 feet.
Electromagnetic Compatibility (EMC, IEC 60601-1-2)
Declaration – electromagnetic emissions
The Philips Avent breast pump is intended for use in the electromagnetic environment specified
below. The user of the Philips Avent breast pump should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The Philips Avent breast pump uses RF energy only for its
internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic
equipment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Class B
The Philips Avent breast pump is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic
emissions IEC
61000-3-2
Class A
Voltage
fluctuations/ flicker
emissions IEC
61000-3-3
Complies
Declaration – electromagnetic immunity
The Philips Avent breast pump is intended for use in the electromagnetic environment specified
below. The user of the Philips Avent breast pump should assure that it is used in such an environment.
English
24
Summary of Contents for AVENT SCD223
Page 1: ...SCF334 SCF332 SCD223 SCD292 ...
Page 2: ......
Page 3: ...SCF334 Fig A SCF332 1 2 4 5 14 15 16 17 18 13 7 6 8 10 11 12 3 9 1 ...
Page 5: ...English 6 Čeština 31 Magyar 56 Polski 83 Română 109 Slovensky 135 Ελληνικα 161 Български 186 ...
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ...28 32 1 2 36 30 34 38 27 31 35 29 33 37 23 24 25 26 ...
Page 214: ......
Page 215: ......
Page 216: ...4213 354 4007 1 www philips com 75 recycled paper ...