Testing the Receiver
Chapter 9 - Maintenance and Performance Assurance
55
Testing the Interface Between the Fetal Monitor and Receiver
1. Remove the fetal monitor interface cable from the back of the
receiver.
2. All the lights on the front of the receiver will go out.
If they are still lit the receiver is not working properly.
Contact your Philips Medical Response Centre.
Testing the Interface Cable
1. Connect the interface cable to the receiver.
2. Remove the interface cable from the fetal monitor.
a. If all the lights on the receiver are lit, the interface cable is
faulty.
b. If the lights on the receiver go out, the telemetry interface in
the fetal monitor is faulty.
Contact your Philips Medical Response Centre.