Solving Common Monitoring Problems
Chapter 4 - Monitoring
33
Solving Common Monitoring Problems
Problem
Possible Causes
Solutions
Monitoring FHR using Ultrasound
Erratic trace.
Patient walking heavily.
Ask patient to walk quietly.
Series 50 Fetal Monitor
Users Only
Suspicious Fetal Movement
Profile.
Maternal movement due
to patient ambulating.
Turn off FMP.
External Monitoring Uterine Activity (Toco) - External
Artifact on Toco trace.
Walk frequency is being
recorded.
None.
Monitoring Intrauterine Pressure (IUP) - Internal
Err 3
displayed when using
the Series 50 T with an old
8040A Monitor.
IUP sensitivity setting is
incorrect.
Contact your Philips Medical
Response Center to change
setting. Once the setting is
changed, you will only be able
to monitor IUP with the Series
50T.