For hygienic reasons and in order to keep
your nebulizer in good working order you
should replace the following parts temporally
limited.
Yearly replacement of the nebulizer plus accesso
ries e.g. compressed air tubing. For this purpose
use our YearSet.
It is absolutely important to carefully follow the
instructions for cleaning and disinfection, in order
to guarantee an impeccable function of the product
and a successful therapy.
Material resistance
The nebulizer must not come into touch neither with
aromatic or halogenized hydrocarbons and ethers
nor – if possible – with aldehydes, esters, ketones,
ali phatic hydrocarbons and oxidizing acids. The manu-
facturer cannot be liable for such cases.
Do not use temperatures above 121°C.
Repeated sterilization does not have any influence on
the nebulizers, due to the properties of the material
used; however the manufacturer accepts no respon-
sibility, if the nebulizers are sterilized more than
50 times.
All parts of the nebulizers must be handled with care
(prevention of mechanical damages).