9
PM5A2
4. Connecting the PM5A2
ATTENTION:
For safe operation, follow the procedures detailed below with the device Powered-Off.
Check if there is no difference of potential voltages between the points where the PM5A2shall be connected to.
Also, check the same between those points and the ground.
At the time of the connection and power-on, the equipment automatically enters in the
voltmeter mode and displays the circuit voltage.
The circuit to be tested must be de-energized to avoid interferences in the measurement. The equipment will block
the start of measurement if it detects a voltage greater than 60 V in the circuit.
Connect the red cable security terminal to the equipment
(-V) output terminal
, the BNC terminal of the black cable
to the
zero reference (+R)
terminal and the “alligator” terminals to the element to be measured as indicated in the
figure.
The test leads in the picture are merely illustrative.