Phenix Technologies PM15-2 Operation And Application Manual Download Page 14

3-4 

SET UP AND OPERATION INSTRUCTIONS 

 

NOTES CONCERNING THE USE GUARD AND GROUND

 

 
 

NORMAL MODE CONFIGURATION

 

In this configuration the 

Return

 and 

Ground

 is connected together with an external jumper or a 

switch internal to the Test Set, depending on the design.  Nothing is connected to the 

Guard

.  

Both stray leakage to earth ground and

 

leakage through the Test Specimen is indicated

 

on the 

current meter.  This configuration should be used if it is not possible to isolate the return side of 
the Test Specimen from earth ground.

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

HIGH VOLTAGE 
TRANSFORMER

 

Amp

 

Guard 

Ground 

Return 

Ground 

Test 

Specimen 

Stray 

Leakage 

High 
Side 

Return or Low 
Side 

Summary of Contents for PM15-2

Page 1: ...OPERATIONS AND APPLICATIONS MANUAL MODEL NUMBER PM15 2 PM15 4 15 kV DC HIGH POTENTIAL TESTER WITH MEGOHMMETER Version 1 0...

Page 2: ......

Page 3: ...TION NOTES DESCRIPTION SPECIFICATIONS CONTROLS AND CONNECTIONS 2 FRONT PANEL CONTROLS AND METERS INPUT OUTPUT CONNECTIONS SET UP AND OPERATION INSTRUCTIONS 3 INSTALLATION TEST LEAD CONNECTIONS OPERATI...

Page 4: ...Equipment have been listed below for your reference Become familiar with your instrument before performing an actual test Know your work area check that all circuits are de energized and locked out N...

Page 5: ...an assistant operator where appropriate to keep non essential personnel away from the test site 7 When testing cables the conductors at the distant end should be isolated from each other and taped 8 W...

Page 6: ...ers Input 120 Volts AC 50 60 Hz single phase 1 Amp or 220 Volts AC 50 60 Hz single phase 5 Amps Output Regulated 0 15 kV DC continuously adjustable with manual control 0 2mA PM15 2 0 4mA PM15 4 Negati...

Page 7: ......

Page 8: ...will sound until voltage control is returned to zero Currentmeter Megohmmeter The currentmeter range is selected by the five position rotary switch The switch setting indicates the full scale current...

Page 9: ...ltage Current Trip This control located below the currentmeter range select switch allows the operator to set the level of leakage current measured by the instrument before high voltage shutdown occur...

Page 10: ...order to measure currents to return in guard mode Ground This terminal is to be connected to the facility ground black cable provided Leakage current or insulation resistance values associated with t...

Page 11: ...connect the test leads 6 Connect the AC Line cord to the power module and to the appropriate AC Power Source see unit specification tag either 120V or 220V TEST LEAD CONNECTIONS An understanding of th...

Page 12: ...ate and remain recessed NOTE If the Voltage Control has not been set to Zero the overcurrent alarm will sound To silence this alarm insure that the voltage control is in its full counter clockwise pos...

Page 13: ...click into position 6 Observe that the Kilovolt meter indicates zero 0 output 7 Turn off the high voltage power supply by depressing the High Voltage power switch Observe that this switch is no longer...

Page 14: ...to the Test Set depending on the design Nothing is connected to the Guard Both stray leakage to earth ground and leakage through the Test Specimen is indicated on the current meter This configuration...

Page 15: ...on the current meter Any current associated with the Guard bypasses the current meter Leakage current through the test specimen is indicated on the current meter Stray leakage to earth ground is not i...

Page 16: ...4 1 SCHEMATIC Drawing Number Description 9109080 PM15 2...

Page 17: ...r Supply 15kV 2mA PM15 2 Ultravolt 15A24 N30 E T0B6 1 1590174 H V POWER SUPPLY Power Supply 15kV 4mA PM15 4 Ultravolt 15C24 N60 E T0B6 ELECTRICAL PARTS COMPONENTS 1 1152590 Module Power Entry Corcom 6...

Page 18: ...2 3306 5 1152260 CON2 3 6 7 9 Connector 4 Pin Red Allied 512 3239 5 1152261 CON2 3 6 7 9 Connector Board 4 Pin Allied 512 3302 1 1152230 CON8 Connector 12 Pin Red AMP 1 640440 2 1 1152231 CON8 Connect...

Page 19: ...istor 110 k 25W 1 1 1722080 R29 Resistor 2 2 k 5w 12 2 1722600 R31 38 Resistor 10 k 25W 1 1 1720180 R34 Resistor 100 25W 1 1 1894335 T1 Transformer ST4 24 Newark 01F027 5 1356300 TP1 5 Test Points Min...

Page 20: ...g parts to include the serial number of the unit as well as the part number of the replacement part When your purchase order is received at our office a representative of Phenix Technologies will cont...

Page 21: ...for Return Cause of Defect If Phenix Technologies Inc deems return of the part appropriate it will then issue an Authorization for Return If return is not deemed advisable other inspection arrangement...

Reviews: