5-1 

 

SECTION 5:  OPERATING INSTRUCTIONS 

 
 

 
 

1) DIVIDER WITH METER MODULE OPERATION: 

 

 

Make sure connections have been made as described in “Set-Up” section. 
 

 

Switch control power switch to 

ON

 position. 

 

 

Select the desired range for the voltmeter based on the expected input signal level. 
 

 

Select desired mode for voltmeter with function switch. 

 

Note:    If the Ripple voltage measurement mode is to be selected, the Range selector must 

be set for the required input voltage level. If ripple voltage is to be read on input 
voltage over 30kV the range selector has to be on high range, regardless of the 
expected ripple voltage level. To avoid damage to the unit ascertain the unit will 
not over-range on the AVG or the DC modes before attempting to read the ripple 
voltages. DO NOT attempt to read ripple voltage on a range lower than required to 
measure the full input voltage. 
 

 

Activate high voltage source. 
 

 

Measure voltage as required.  Meter low range should not be used past 30kV.  Meter may 
not be accurate past 100% of reading and actual voltage may be much higher than 
indicated. 
 

 

De-activate high voltage source and assure, as appropriate, that high voltage source is de-
energized and discharged.   
 

 

When testing is complete, switch control power to 

OFF

 position. 

 

 

After assuring that high voltage source and unit are de-energized and discharged, all leads 
and connections may be disconnected and unit packaged for transport or storage.   

 

 

Do not subject unit to Flashovers. Damage may occur.  If meter becomes disabled from 
Flashover or Transient condition, refer to Troubleshooting section.  

 

 

Battery operational time will be extended if Backlight is kept at minimum setting.  

 

 

Low battery lamp comes on at approximately 20% of charge.  If unit is run too long after low 
battery lamp comes on, it will shut down and must be recharged before it will turn on again.  

WARNING 

THIS UNIT SHOULD ONLY BE OPERATED BY PERSONS KNOWLEDGEABLE OF HIGH 

VOLTAGE TESTING AND SAFETY PROCEDURES.  IMPROPER OPERATION 
MAY RESULT IN INJURY OR DEATH.   

Summary of Contents for KVM300

Page 1: ...KVM300 AC KILOVOLTMETER MODEL NO KVM300 VERSION 2 1 Copyright Phenix Technologies Inc JPG SLU MAY 2008 REV 9 4 14 nab...

Page 2: ...ECTRICAL SET UP OPERATIONAL NOTES 4 OPERATING INSTRUCTIONS 5 CALIBRATION 6 MECHANICAL MAINTENANCE 7 STORAGE OF EQUIPMENT 8 TROUBLESHOOTING 9 RECOMMENDED SPARE PARTS 10 PARTS ORDERING INFORMATION 11 RE...

Page 3: ...KVM300 1 1 SECTION 1 DANGER WARNINGS WARNING Complete Grounding of this unit is necessary for the safe operation of this equipment Disconnect inputs before ungrounding this equipment...

Page 4: ...h Voltage Test Equipment have been listed below for your reference Become familiar with your instrument before performing an actual test Know your work area check that all circuits are de energized an...

Page 5: ...try Display is 4 Digit LED with adjustable backlighting ACCURACY 1 of Reading 3 counts from 10 100 of Range FUNCTION SELECTIONS 1 Peak 2 Peak responding calibrated to RMS Value 2 Peak Peak responding...

Page 6: ...round 6 BNC INPUT CONNECTOR This BNC input connector connects the meter module to the divider 7 BACKLIGHT DIMMER This dial adjusts the intensity of the backlight for use during low light conditions 8...

Page 7: ...inspection to determine if the unit was damaged in shipment If there are any signs of physical damage such as dents scratches or oil leaks contact the factory before proceeding NOTE Read and understa...

Page 8: ...will affect the AC calibration CAUTION The meter module has to be grounded for safe and proper operation Connect the ground post on the metering module to an appropriate ground A 10 lead is supplied...

Page 9: ...sure voltage as required Meter low range should not be used past 30kV Meter may not be accurate past 100 of reading and actual voltage may be much higher than indicated De activate high voltage source...

Page 10: ...and 1 M 200pF AC for the highest possible accuracy Readings will correspond to the type of meter used AVG will show AVG RMS will show RMS etc Use DC range for DC signals 10M input meter Select desire...

Page 11: ...TION SHOULD ONLY BE DONE BY PERSONS FAMILIAR WITH HIGH VOLTAGE TESTING AND SAFETY PROCEDURES CALIBRATION PROCEDURES Due to the complexity and the standards required to calibrate this instrument Phenix...

Page 12: ...ainst the elements in normal use It is recommended that the finish be wiped down with an ordinary household cleaner for longer life and for proper electrical operation of the unit Also inspect all fab...

Page 13: ...olonged period the following precautions are recommended The equipment should be covered and kept in a warm dry environment 95 maximum humidity 5 to 50 degrees C Prior to placing the equipment back in...

Page 14: ...g testing because of transient discharge occurrence and won t come back on by cycling the power switch Carefully remove front panel and unplug the battery pack for approximately 30 seconds then retry...

Page 15: ...r quantity of parts should be stocked as spares In such a case contact your Phenix Technologies sales representative for a recommendation Current prices may be obtained by contacting the Parts Orderin...

Page 16: ...important when ordering parts to include the serial number of the unit as well as the part number of the replacement part When your purchase order is received at our office a representative of Phenix...

Page 17: ...er Reason for Return Cause of Defect If Phenix Technologies Inc deems return of the part appropriate it will then issue an Authorization for Return If return is not deemed advisable other inspection a...

Page 18: ...D TR M K K K TB X Intensite Diode Transisteur Thyristor Borne Ouvert Ferme Contact de SYMBOLE ZU SCHEMA SYMBOLES POUR SCHEMA DE CIRCUIT CIRCUIT DIAGRAM SYMBOLS Variable Resisitor Resistor Lamp Indica...

Page 19: ...KVM300 14 1 SECTION 14 SCHEMATICS AND DRAWINGS Drawing Number Description 1 7916004 KVM300 SCHEMATIC...

Page 20: ...or s Manual are correct If there are concerns or comments as you have used this information Phenix Technologies appreciates any feedback Unit Serial Number Sect Page s Comment Please return to Phenix...

Reviews: