HC-20C
1-2
SPECIFICATIONS
Instrumentation:
Currentmeter:
Resolution: 0.1% of Full Scale
Ranges: 100/1000/10,000/100,000 Amps
Accuracy: 0.8% rdg + 0.3% range
± 5% of Full Scale 0.02
– 0.10 Seconds
Timer:
Resolution: 0.002 Seconds
0.120 Cycles
Ranges: 0 - 999.999 Seconds
0 - 99,999.990 Cycles
Test durations of less than 10 cycles
Timer Accuracy: ± 0.004 Seconds
± 0.250 Cycles
Test durations of greater then 10 cycles
Timer Accuracy: ± 0.010 Seconds
± 0.600 Cycles