![Phcbi PF-L5181W-PA Operating Instructions Manual Download Page 16](http://html1.mh-extra.com/html/phcbi/pf-l5181w-pa/pf-l5181w-pa_operating-instructions-manual_1536785016.webp)
15
English
START-UP OF UNIT
The following procedures should be adhered to for initial start-up and continuous operation.
1.
Connect the unit to dedicated power supply. Do not put any product in the unit at this time.
2.
Set the desired temperature. The unit is preset to operate at -20°C. This is the optimal storage
temperature to maintain vaccine and biologic products within -15°C to -25°C throughout the usable
storage area.
3.
Confirm that the chamber temperature is at the desired temperature.
4.
When you are satisfied that the unit is working correctly, begin slowly placing product into the chamber to
minimize the temperature rise. Avoid long or excessive door openings to insure stored product integrity.
OPERATING INSTRUCTIONS
Temperature Setting
Table 2 shows the basic operation method. Perform key operation in the sequence indicated in the table. The
example in the table is based on the assumption that the
freezer temperature is -18
o
C.
Note:
The unit is set at the factory at -20
o
C.
Table 2 Basic operation procedure (Example of setting: Freezer @ -19
o
C)
Operation
Key operated Display after the key operation
1
Connect to the power source.
-----
The current freezer temperature
is displayed after 5 seconds of
connecting the unit to the power
source. When freezer
temperature is higher than
-15°C, HI is displayed and LO
is displayed when lower than
–25
C.
2
Press SET key.
SET
The digits of the temperature display flash.
3
Set to -19 by using the down arrow
key.
Pressing and releasing the down
arrow key shifts down the set point
by 1°C increments.
4
Press and release SET key.
SET
The value (-19) is stored in
memory and the current freezer
temperature is displayed.
Note:
If no key has been pressed for 90 seconds in the temperature set mode, the display mode returns automatically
to the temperature display mode. In this case, the chamber temperature setting is not changed. The freezer
temperature can be set in the range between -15°C and -25°C. The guaranteed temperature with no load at an
ambient temperature of 22°C is -15 to -25°C.