![Phcbi PF-L5181W-PA Operating Instructions Manual Download Page 13](http://html1.mh-extra.com/html/phcbi/pf-l5181w-pa/pf-l5181w-pa_operating-instructions-manual_1536785013.webp)
12
English
CONTROL PANEL FUNCTIONALITY CHECK
The first time the unit is powered on, you must insure the control panel functionality by following the operation
described in table 1:
Table 1 Manual check of Control panel
Operation
Key operated
Control panel description
1
Connect to the power
source.
-----
Five seconds after turning on the unit, the temperature inside the
unit should appear on temperature display (LED Screen).
2
Press “Alarm Test”
key.
ALARM TEST
Two seconds after pressing the button, the buzzer should sound
every 1 second. Let the buzzer sound at least 3 times.
3
Press “Alarm Test”
key.
ALARM TEST
Within 2 second of pressing the button, the buzzer should stop.
(The buzzer should stop sounding after two second of pressing
the button).
4
Press “SET” key.
SET
Within 2 seconds of pressing the button, you will hear a beep
once
. The display will show “-20” on LED Screen and it will start
blinking.
5
Press “
△
” key.
Within 2 seconds of pressing the button, you will hear a beep
once. The display will show “-19” on LED Screen and it will start
blinking.
6
Press “
△
” key two
times.
Within 2 seconds of pressing the button, you will hear a beep
once. The display will show “-17” on LED Screen and it will start
blinking.
7
Pres
s “
▽
” key three
times.
Within 2 seconds of pressing the button, you will hear a beep
once. The display will show “-20” on LED Screen and it will start
blinking.
9
Open the door
-----
Two seconds after opening the door, the door check lamp will
turn red.
10
Close the door
-----
Two seconds after closing the door, the door check lamp will turn
off.
11
Open the door
-----
Two seconds after opening the door, the door check lamp will
turn red.
12
Keep the door open
-----
After 120 seconds, the alarm lamp will turn red and will start
beeping every second. Let the unit beep at least 3 times.
13
Close the door
-----
Two seconds after closing the door, the door check lamp will turn
off.
“
Summary of Contents for PF-L5181W-PA
Page 25: ...24 English SERVICE Wiring Diagram...
Page 27: ...26 English MEMO...