15
KEY LOCK FUNCTION
The temperature setting of chamber can be locked to avoid accidental change. When the lock is ON,
the change of chamber temperature is unable even if the key on the control panel is operated.
Initial setting (at factory): Lock is OFF
Display Mode
Function
L 0
Lock is OFF
Enable to change of temperature setting
L 1
Lock is ON
Disable to change of temperature setting
Example: Changing the lock status to ON from OFF (initial setting).
Description Operation
Key operated
Indication after operation
1 -----
-----
The current chamber temperature
is displayed.
2
Press the digit shift key for about 5
seconds.
The current setting is displayed and
the first digit blinks.
3
Press the numerical value shift
key and scroll the figure to 1
When pressed, the figure of settable
digit changes.
4 Press the set key
SET
The key lock is set to ON.
The current chamber temperature
is displayed.
■
The setting mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.
DEFROSTING
The following 2 types of defrosting methods is provided for the refrigerator. Both of them are controlled
automatically.
●
Cycle defrosting
The chamber temperature is kept stable by ON/OFF operation of the compressor. When the
compressor is OFF (stopped), the frost on the evaporator is melted by the heater. This defrosting never
influences to the chamber temperature.
●
Defrosting by defrost sensor
The cycle defrosting may not be sufficient to remove the frost on the evaporator when the ambient
temperature is high, the door is opened frequently, or heavy moist items are stored in the chamber. In
this case, the defrosting operation is started automatically when the defrost sensor detects the frost.
During the defrosting, the current chamber temperature and dF is displayed alternately on the
temperature display.
After completion of defrosting, dF display is disappeared and normal operation resumes.
<Important>
■
During the defrosting, the chamber temperature reaches around 10
o
C temporary.
■
Too much frost is accumulated on the evaporator when the chamber is running under high temperature
and humidity condition. For example, the unit starts to defrost once a week with 2
o
C setting in the
ambient of 30
o
C and 80%R.H.
Summary of Contents for MPR-S163 Series