26
SETTING OF COMPRESSOR START DELAY
The delay time of compressor start can be changed to reduce the load on the power line and to facilitate
the start-up (reset) of the freezer at the recovery after power failure.
Setting range of delay time: between 3 and 15 minutes with 1 minute interval
Initial setting (factory setting): 3 minutes
Note:
There is no need of changing the delay time when the capacity of power source is adequate.
Example: Change the delay time to 4 minutes from 3 minutes
►
Following is an example. Operate the key according to the desired delay time.
Description of operation
Key operated
Indication after operation
1
-----
The current chamber temperature is
displayed.
2
Depress the up arrow key for 5
seconds.
F00 is displayed and the first figure
0 blinks.
3
Depress the up arrow key and
scroll the figure to 5.
The display changes to F05 from
F00.
4 Depress the set key.
SET
The current setting (003) is displayed
and the first figure 3 blinks.
5
Depress the up arrow key and
scroll the figure to 4.
The display changes to 004 from
003.
6 Depress the set key.
SET
The delay time is memorized and the
current chamber temperature is
displayed.
The setting mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation (auto-return function). In this case, the setting is not accepted.
<Important>
Never input any function code other than the specified one.