SETTING OF BUZZER SUSPENDED PERIOD
The abnormality is noticed again by sounding of the buzzer after certain suspended period (buzzer
recovery timing) if the alarm status continues even after the buzzer is silenced by depressing the buzzer
stop key (BUZZER). Always set the buzzer suspended period to avoid the wrong recognition of alarm
status
Setting range of buzzer suspended period:
Between 10 and 60 minutes with 10 minutes interval (display of setting; between 010
and 060) or no recovery (display of setting; 000)
Initial setting (factory setting): 30 minutes (display of setting; 030)
The buzzer does not recover once the buzzer is silenced by depressing the buzzer stop key (BUZZER)
when the suspended time is set to 000. However, the buzzer operates if other alarm status is detected.
Example: Change the suspended period to 20 minutes from 30 minutes
►
Following is an example. Operate the key according to the desired buzzer suspended period.
Description of operation
Key operated
Indication after operation
1
-----
The current chamber temperature is
displayed.
2
Depress the up arrow key for 5
seconds.
F00 is displayed and the first figure
0 blinks.
3
Depress the up arrow key and
scroll the figure to 5.
The display changes to F05 from
F00.
4 Depress the scroll key once.
The second figure 0 blinks.
5
Depress the up arrow key and
scroll the figure to 2.
The display changes to F25 from
F05.
6 Depress the set key.
SET
The current setting (030) is displayed
and the second figure 3 blinks.
7
Depress the up arrow key and
scroll the figure to 2.
The display changes to 020 from
030.
8 Depress the set key.
SET
The suspended period is memorized
And the current chamber temperature
Is displayed.
The setting mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation (auto-return function). In this case, the setting is not accepted.
The setting should be performed during normal operation, not during alarm status.
<Important>
Never input any function code other than the specified one.