31
Filling water reservoir
To fill the water reservoir, do the following:
1.
Open the outer door and remove the lowest tray.
See Fig.1.
2.
Remove a screw in front side and rotate the water
supply inlet cover to open the water supply inlet.
See Fig.2.
3.
When there is no water in the water reservoir (for
example, first time to operate), add approximately 20
L of preheated sterile distilled water at 37
o
C. When
the water level alarm lamp (RH PAN) is blinking, add
approximately 15 L of preheated sterile distilled
water at 37
o
C. See Fig.3. It might take
approximately 2 minutes until the lamp goes off.
4.
Make sure the water level alarm lamp (RH PAN)
on the control panel does not blink.
5.
Rotate back the water supply inlet cover and
tighten the removed screw.
6.
Put the lowest tray back in place and close the
outer door.
Note:
• Add sterile distilled water as water for humidifying.
The ion exchange water is not fit to use.
• The sterile water filled in the water reservoir should
be pre-heated at 37
o
C. Cold water lowers the
chamber temperature and humidity.
Fig. 1
Fig. 2
Fig. 3
Water supply inlet