   START-UP OF UNIT 

 

The following procedures should be adhered to for initial start-up and continuous operation 

 

1.

 After enough ventilation, install the shelves.    Use the stainless steel plate for the bottom shelf. 

 

2.

 Leave the alarm and buzzer switches off and set the temperature recorder. 

 

3.

 Connect unit to dedicated power supply and turn on the power switch.    Do not put any product in the 

unit at this time. 
 

4.

 Allow the chamber temperature to fall to +4 

o

C.    Check the chamber temperature by the indication of 

the digital thermometer. 
 

5.

 When the chamber temperature drops to +4 

o

C, turn on the alarm and buzzer switches.    Make sure 

that the automatic temperature recorder indicates the temperature correctly. 
 

6.

 Press the alarm test switch to check the warning device functions properly. 

 

7.

 Turn on the fluorescent light switch to check the light. 

 

8.

 When you are satisfied that the unit is working correctly, begin slowly placing product into the chamber 

to minimize the temperature rise. 
 

NOTICE 

 

Door alarm by opening and closing the door 

 

1.

 When the door is opened and remains open for less than 3 minutes and items are removed or replaced 

gradually, the alarm will not be activated.    But if the door is frequently opened or remains open for a long 
time or many items are put in at the same time, the alarm will be activated.  The alarm may also be 
activated after replacing items and closing the door.    This does not denote a malfunction.    It is caused 
by the circulation fan stopping while the door is open and starting when the door is closed. 
 

2.

 When the alarm is activated, run the unit until the temperature indicator needle falls below the upper 

limit (generally it takes 10 to 30 minutes).    Reset by turning the alarm switch off and turning it on again.   
The buzzer switch can be turned off while you are waiting for the temperature drop.    After the chamber 
temperature falls to adequate value, be sure to turn on the buzzer switch again. 

 

   

Adjustment of shelves

 

 
The shelves can be adjusted to the height of items place 
inside.  To adjust the height, insert the shelf support into 
the slots at the desired height. 
 
 
 
 
 

Summary of Contents for MBR-107D

Page 1: ...ries Please read the operating instructions carefully before using this product and keep the operating instructions for future use See page 29 for all model numbers Operating Instructions Blood Bank R...

Page 2: ...15 OPERATION CONTROL P 16 REMOTE ALARM TERMINAL P 16 TEMPERATURE RECORDER Usage of temperature recorder P 17 Setting of temperature recorder P 17 ROUTINE MAINTENANCE Cleaning of cabinet P 19 Cleaning...

Page 3: ...d to change without notice due to the improvement of performance or functions Contact our sales representative or agent if any page of the operating instructions is lost or page order is incorrect Con...

Page 4: ...t in a hazard to personnel possibly resulting in serious injury or death CAUTION Failure to observe CAUTION signs could result in injury to personnel and damage to the unit and associated property Sym...

Page 5: ...wer supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may...

Page 6: ...al operation may cause electric shock or fire When removing the plug from the power supply outlet grip the power supply plug not the cord Pulling the cord may result in electric shock or fire by short...

Page 7: ...turning off of power switch The stored items may be damaged due to the change of setting Be careful not to tip over the unit during movement to prevent damage or injury Prepare a safety check sheet wh...

Page 8: ...8 REFRIGERATOR COMPONENTS 1 22 21 20 19 17 16 18 15 14 13 12 2 3 4 5 6 7 8 9 10 11...

Page 9: ...echanism Magnet door seal ensures a tight seal being formed when the door is closed 9 Temperature recorder 10 Kick plate When the unit is placed close to the wall the kick plate functions as an air ou...

Page 10: ...cell battery A roll of recording paper can be used for one month Temperature from 20 o C to 6 o C is recorded in increment of 1 o C Since it is a pressure sensitive type there is no need to supply in...

Page 11: ...r is open 8 Digital thermometer This thermometer indicates temperature of air within refrigerator It is integrated with temperature regulating circuit Since in this refrigerator cooling and heating ar...

Page 12: ...e condition may cause vibration or noise WARNING Install the unit on a sturdy floor If the floor is not strong enough or the installation site is not adequate this may result in injury from the unit f...

Page 13: ...eveling feet Extend the leveling feet by rotating them counterclockwise to contact them to the floor Ensure the unit is level See Fig 1 3 Fixing the unit Two hooks are attached to the rear of the cabi...

Page 14: ...cable can be led through the access port on the cabinet After installation a rubber cap should be used to seal the access port Failure to do this can affect the temperature uniformity inside the cabin...

Page 15: ...to minimize the temperature rise NOTICE Door alarm by opening and closing the door 1 When the door is opened and remains open for less than 3 minutes and items are removed or replaced gradually the a...

Page 16: ...sure to set the alarm switch to on again Protective circuit If the temperature controller malfunctions or if the chamber temperature falls or rises the unit is designed so that the chamber temperatur...

Page 17: ...dle is higher than the upper limit or lower than the lower limit the alarm device will be activated Setting of temperature recorder As shown in the Fig 1 pull the knob on the upper part of the tempera...

Page 18: ...g paper outlet If not all of the recording paper has been fed into the used recording paper compartment send all the recording paper in the compartment first turning the gear Battery change To change...

Page 19: ...iluted detergent can damage the plastic components For the dilution refer to the instruction of the detergent After the cleaning with the diluted detergent always wipe it off with a wet cloth Then wip...

Page 20: ...e the two screws below the shade 2 The fluorescent light can be removed by gently pulling it down 3 Insert a new fluorescent light bulb after removing the old one from the socket 4 Be sure to insert t...

Page 21: ...rm test switch is pressed The nickel cadmium batteries are not completely charged If so run the unit for about 3 hours and check again If only the buzzer or only the alarm lamp functions when the alar...

Page 22: ...el by using a screw driver 3 Pull out the control panel and then remove 2 screws mounting the warning device to the panel see Fig 3 4 Detach 2 connectors from the warning device Remove 4 screws fixing...

Page 23: ...ersonnel Always remove doors to prevent accidents such as suffocation Recycle of battery The unit contains a rechargeable battery The battery is recyclable At the end of its useful life check with you...

Page 24: ...the chemical involved German Entsorgung von Altger ten und Batterien Nur f r die Europ ische Union und L nder mit Recyclingsystemen Dieses Symbol auf den Produkten der Verpackung und oder den Begleitd...

Page 25: ...r sur les piles pictogramme du bas Si ce pictogramme est combin avec un symbole chimique il r pond galement aux exigences pos es par la Directive relative au produit chimique concern Spanish Eliminaci...

Page 26: ...Neste caso estar em conformidade com o estabelecido na Diretiva referente aos produtos qu micos em causa Italian Smaltimento di vecchie apparecchiature e batterie usate Solo per Unione Europea e Nazi...

Page 27: ...l Onderstaand symbool Dit symbool kan in combinatie met een chemisch symbool gebruikt worden In dit geval volstaan de eisen die zijn vastgesteld in de richtlijnen van de desbetreffende chemische stof...

Page 28: ...r Fin and tube type Forced air circulation Refrigerant R 134a Defrosting Forced type Fully automatic Heater for chamber 20 W Temperature control Microprocessor control system Temperature alarm Nickel...

Page 29: ...er accuracy 1 o C 2 o C to 8 o C Alarm duration 9 hours for power failure Noise level 50 dB A scale Maximum pressure 1569 kPa Rated voltage alternate current 110 V 220 V 220 V 230 V 240 V 220 V Rated...

Page 30: ...Contamination of the unit Unit interior No contamination Yes No Decontaminated Yes No Contaminated Yes No Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to...

Page 31: ......

Page 32: ...Printed in Japan LDCL037400 0 S0418 0 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan PHC Corporation 2018...

Reviews: