6
Maintenance
6
Maintenance
6.1 Replacing the filling of the air drier
Change zeolite, if the saturation level is reached (the orange colored dry beads are
bleached) .
Fig. 2:
Changing the zeolite filling
Dismantle the air drier TTV 001 by unscrewing the knurled screws 11.
Unscrew the upper heagon nut 12 and remove flat seal 8.
Hold together glass cylinder 7 and the lower flange 1.
Remove upper felt washer 4.
Remove old zeolite.
Hold together glass cylinder 7 and the lower flange 1.
Ensure correct position of the lower flat seal 9 and the lower felt waher 4 (correct po-
sition as shown).
Fill in new zeolite (total replacement filling, shake the zeolite together by carefully tap-
ping on the glass cylinder).
Insert upper felt washer 4.
Check correct position of the upper flat seal 9.
Place upper flange 1 onto glass cylinder 7.
Place flat seal 8 over threaded rod 3 onto flange 1.
Screw on hexagon nut 12.
Refit air drier TTV 001 with knurled screws 11.
1
Flange
2
Hose nipple
3
Threaded rod
4
Felt washer
5
Valve spindle
7
Glass cylinder
8
Flat seal
9
Flat seal
11
Knurled screw
12
Hexagon nut
4
2
12
4
3
8 1
11
9
9
11
1
7
5